Improved precision of SPECT myocardial blood flow using a net tracer retention model

Sarah G Cuddy-Walsh iD, Robert A deKemp iD, Terrence D Ruddy iD, R Glenn Wells iD

Medical Physics · 2023

DOI: 10.1002/mp.16186 ↗

At a glance

Journal article in Medical Physics, 2023. Written by Sarah G Cuddy-Walsh and 3 co-authors. Institution: Division of Cardiology, University of Ottawa Heart Institute, Ottawa, Ontario, Canada. Major MEDLINE topics: Myocardial Perfusion Imaging, Coronary Artery Disease. Cited 8 times so far (Crossref); 28 references. Funded by: GE Healthcare and NSERC.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

Noninvasive quantification of absolute myocardial blood flow (MBF) and myocardial flow reserve (MFR) provides incremental benefit to relative myocardial perfusion imaging (MPI) to diagnose and manage heart disease. MBF can be measured with single‐photon emission computed tomography (SPECT) but the uncertainty in the measured values is high. Standardization and optimization of protocols for SPECT MBF measurements will improve the consistency of this technique. One element of the processing protocol is the choice of kinetic model used to analyze the dynamic image series.

Purpose

This study evaluates if a net tracer retention model (RET) will provide a better fit to the acquired data and greater test‐retest precision than a one‐compartment model (1CM) for SPECT MBF, with (+MC) and without (‐MC) manual motion correction.

Methods

Data from previously acquired rest‐stress MBF studies (31 SPECT‐PET and 30 SPECT‐SPECT) were reprocessed ± MC. Rate constants (K1) were extracted using 1CM and RET, +/‐MC, and compared pairwise with standard PET MBF measurements using cross‐validation to obtain calibration parameters for converting SPECT rate constants to MBF and to assess the goodness‐of‐fit of the calibration curves. Precision (coefficient of variation of test re‐test relative differences, COV) of flow measurements was computed for 1CM and RET ± MC using data from the repeated SPECT MBF studies.

Results

Both the RET model and MC improved the goodness‐of‐fit of the SPECT MBF calibration curves to PET. All models produced minimal bias compared with PET (mean bias < 0.6%). The SPECT‐SPECT MBF COV significantly improved from 34% (1CM+MC) to 28% (RET+MC, P = 0.008).

Conclusion

The RET+MC model provides a better calibration of SPECT to PET and blood flow measurements with better precision than the 1CM, without loss of accuracy.

Subjects

MeSH terms (MEDLINE)

Myocardial Perfusion ImagingCoronary Artery DiseaseHumansCoronary CirculationTomography, Emission-Computed, Single-PhotonRadiopharmaceuticalsMyocardiumSingle Photon Emission Computed Tomography Computed TomographyPositron-Emission Tomography

Author keywords

SPECTmyocardial blood flow (MBF)myocardial flow reserve (MFR)net tracer retention modelprecisiontissue compartment model

Substances

Radiopharmaceuticals

Authors and institutions

  • Division of Cardiology, University of Ottawa Heart Institute, Ottawa, Ontario, Canada.
  • Division of Cardiology, University of Ottawa, Ottawa, Ontario, Canada.

ORCID profiles: Sarah G Cuddy-Walsh, Robert A deKemp, Terrence D Ruddy, R Glenn Wells

Funding

  • GE Healthcare and NSERC

Cite

APA

Cuddy-Walsh, S. G., deKemp, R. A., Ruddy, T. D., & Wells, R. G. (2023). Improved precision of SPECT myocardial blood flow using a net tracer retention model. Medical Physics. https://doi.org/10.1002/mp.16186

Vancouver

Cuddy-Walsh SG, deKemp RA, Ruddy TD, Wells RG. Improved precision of SPECT myocardial blood flow using a net tracer retention model. Medical Physics. 2023. doi:10.1002/mp.16186

BibTeX
@article{CuddyWalsh2023,
  author = {Cuddy-Walsh, Sarah G and deKemp, Robert A and Ruddy, Terrence D and Wells, R Glenn},
  title = {Improved precision of SPECT myocardial blood flow using a net tracer retention model},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16186}
}
Start for free →