Efficacy of ultrasound vector flow imaging in tracking omnidirectional pulsatile flow
Medical Physics · 2023-03 · 50 (3) · 1699-1714
DOI: 10.1002/mp.16168 ↗At a glance
Journal article in Medical Physics, published online on 2023-01-14. Written by Jonathas Haniel and 4 co-authors. Institution: Schlegel Research Institute for Aging and Department of Electrical & Computer Engineering, University of Waterloo, Waterloo, Ontario, Canada. Major MEDLINE topics: Pulse Wave Analysis, Carotid Arteries. Cited 8 times so far (Crossref); 49 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Background
Ultrasound vector flow imaging (VFI) shows potential as an emerging non‐invasive modality for time‐resolved flow mapping. However, its efficacy in tracking multidirectional pulsatile flow with temporal resolvability has not yet been systematically evaluated because of the lack of an appropriate test protocol.
Purpose
We present the first systematic performance investigation of VFI in tracking pulsatile flow in a meticulously designed scenario with time‐varying, omnidirectional flow fields (with flow angles from 0° to 360°).
Methods
Ultrasound VFI was performed on a three‐loop spiral flow phantom (4 mm diameter; 5 mm pitch) that was configured to operate under pulsatile flow conditions (10 ml/s peak flow rate; 1 Hz pulse rate; carotid pulse shape). The spiral lumen geometry was designed to simulate recirculatory flow dynamics observed in the heart and in curvy blood vessel segments such as the carotid bulb. The imaging sequence was based on steered plane wave pulsing (−10°, 0°, +10° steering angles; 5 MHz imaging frequency; 3.3 kHz interleaved pulse repetition frequency). VFI's pulsatile flow estimation performance and its ability to detect secondary flow were comparatively assessed against flow fields derived from computational fluid dynamics (CFD) simulations that included consideration of fluid‐structure interactions (FSI). The mean percentage error (MPE) and the coefficient of determination ( R 2 ) were computed to assess the correspondence of the velocity estimates derived from VFI and CFD‐FSI simulations. In addition, VFI's efficacy in tracking pulse waves was analyzed with respect to pressure transducer measurements made at the phantom's inlet and outlet.
Results
Pulsatile flow patterns rendered by VFI agreed with the flow profiles computed from CFD‐FSI simulations (average MPE: –5.3%). The shape of the VFI‐measured velocity magnitude profile generally matched the inlet flow profile. High correlation exists between VFI measurements and simulated flow vectors (lateral velocity: R 2 = 0.8; axial velocity R 2 = 0.89; beam‐flow angle: R 2 = 0.98; p < 0.0001 for all three quantities). VFI was found to be capable of consistently tracking secondary flow. It also yielded pulse wave velocity (PWV) estimates (5.72 ± 1.02 m/s) that, on average, are within 6.4% of those obtained from pressure transducer measurements (6.11 ± 1.15 m/s).
Conclusion
VFI can consistently track omnidirectional pulsatile flow on a time‐resolved basis. This systematic investigation serves well as a quality assurance test of VFI.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Schlegel Research Institute for Aging and Department of Electrical & Computer Engineering, University of Waterloo, Waterloo, Ontario, Canada.
- Department of Mechanical Engineering, Universidade Federal de Minas Gerais, Belo Horizonte, Minas Gerais, Brazil.
Cite
APA
Haniel, J., Yiu, B. Y. S., Chee, A. J. Y., Huebner, R., & Yu, A. C. H. (2023). Efficacy of ultrasound vector flow imaging in tracking omnidirectional pulsatile flow. Medical Physics, 50(3), 1699-1714. https://doi.org/10.1002/mp.16168
Vancouver
Haniel J, Yiu BYS, Chee AJY, Huebner R, Yu ACH. Efficacy of ultrasound vector flow imaging in tracking omnidirectional pulsatile flow. Medical Physics. 2023;50(3):1699-1714. doi:10.1002/mp.16168
BibTeX
@article{Haniel2023,
author = {Haniel, Jonathas and Yiu, Billy Y S and Chee, Adrian J Y and Huebner, Rudolf and Yu, Alfred C H},
title = {Efficacy of ultrasound vector flow imaging in tracking omnidirectional pulsatile flow},
journal = {Medical Physics},
year = {2023},
volume = {50},
number = {3},
pages = {1699--1714},
doi = {10.1002/mp.16168}
}
EUVIMED