Vergleichsstudie in der Zeitschrift Knee Surgery, Sports Traumatology, Arthroscopy, 2026. Verfasst von Gwo-Chin Lee und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics, Hospital for Special Surgery, New York, New York, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Robotic Surgical Procedures, Knee Joint, Ligaments, Articular. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Robotic total knee arthroplasty (TKA) enables the acquisition of ligament balance data at various stages in the surgical workflow to inform implant planning. This study aims to compare pre‐resection joint balancing versus inverse‐kinematic workflows on implant positioning, joint line obliquity and coronal plane alignment of the knee (CPAK) phenotype.

Methods

A single‐surgeon single‐institution series of 422 primary posterior‐cruciate sacrificing TKAs using a digital ligament tensioner was reviewed. The first 213 cases assessed gaps after tibial and before femoral resections (tibia‐first group) in an inverse‐kinematic alignment (iKA) workflow. The subsequent 174 cases assessed gaps before any bone resections (pre‐resection group) in a femur‐first kinematic alignment (KA) workflow. Native and postoperative medial proximal tibial angle (MPTA) and lateral distal femoral angle (LDFA) were compared using Wilcoxon rank‐sum tests, and CPAK phenotype distributions were compared using chi‐square tests.

Results

Mean preoperative MPTA and LDFA were within 0.5° between groups. Postoperative measurements differed between tibia‐first and pre‐resection for both MPTA (88.3° ± 1.5° vs. 86.7° ± 1.7°, p < 0.001) and LDFA (90.7° ± 2.1° vs. 89.4° ± 2.4°, p < 0.001). The tibia‐first group had greater preoperative to postoperative change in MPTA (0.7° ± 1.3° vs. −0.5° ± 1.7°, p < 0.001), LDFA (2.5° ± 1.7° vs. 1.4° ± 1.5°, p < 0.001) and joint line obliquity (3.2° ± 2.5° vs. 0.9° ± 2.6°, p < 0.001). The pre‐resection group had more patients whose CPAK phenotype did not change postoperatively (35% vs. 25%, p < 0.05), and fewer patients with CPAK changes > 1 (9% vs. 29%, p < 0.001). Final mediolateral balance was within 1 mm of neutral for both groups throughout flexion.

Conclusions

While both techniques achieved a similar balance, a pre‐resection ligament balancing workflow better restored native joint line obliquity and CPAK phenotype compared to an iKA workflow. Level of Evidence Level III.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeRobotic Surgical ProceduresKnee JointLigaments, ArticularHumansBiomechanical PhenomenaFemaleMaleAgedWorkflowMiddle AgedRetrospective StudiesTibiaFemurPosterior Cruciate Ligament

Stichwörter der Autor:innen

inverse‐kinematic alignmentjoint line obliquitykinematic alignmentpre‐resection balancingrobotic‐assisted total knee arthroplasty

Autor:innen und Institutionen

  • Department of Orthopaedics, Hospital for Special Surgery, New York, New York, USA.
  • Corin Group, Clinical Innovation, Raynham, Massachusetts, USA.
  • Viroqua Center for Orthopaedic Surgery, Vernon Memorial Healthcare, Viroqua, Wisconsin, USA.

ORCID-Profile: Gwo-Chin Lee

Zitieren

APA

Lee, G. C., Orsi, A. D., Plaskos, C., & Lawrence, J. M. (2026). Pre‐resection ligament balance more frequently reproduces joint line orientation compared to inverse‐kinematic workflow in robotic total knee arthroplasty. Knee Surgery, Sports Traumatology, Arthroscopy. https://doi.org/10.1002/ksa.70178

Vancouver

Lee GC, Orsi AD, Plaskos C, Lawrence JM. Pre‐resection ligament balance more frequently reproduces joint line orientation compared to inverse‐kinematic workflow in robotic total knee arthroplasty. Knee Surgery, Sports Traumatology, Arthroscopy. 2026. doi:10.1002/ksa.70178

BibTeX
@article{Lee2026,
  author = {Lee, Gwo-Chin and Orsi, Alexander D and Plaskos, Christopher and Lawrence, Jeffrey M},
  title = {Pre‐resection ligament balance more frequently reproduces joint line orientation compared to inverse‐kinematic workflow in robotic total knee arthroplasty},
  journal = {Knee Surgery, Sports Traumatology, Arthroscopy},
  year = {2026},
  doi = {10.1002/ksa.70178}
}
Kostenlos starten →