Optimal intersurface stability for unicompartmental femoral component design with two pegs placed on the distal resection surface: 5 mm peg length increment and 10° peg inclination
Knee Surgery, Sports Traumatology, Arthroscopy · 2024
DOI: 10.1002/ksa.12207 ↗Kurz gefasst
Fachartikel in der Zeitschrift Knee Surgery, Sports Traumatology, Arthroscopy, 2024. Verfasst von Yandong Sun und 9 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering & Med-X Research Institute, Shanghai Jiao Tong University, Shanghai, China. Hauptthemen laut MEDLINE: Prosthesis Design, Finite Element Analysis, Knee Prosthesis, Arthroplasty, Replacement, Knee. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Fundamental Research Funds for the Central Universities, Science and Technology Innovation Action Plan' of Science and Technology Commission of Shanghai Municipality.
Abstract (Originalsprache)
Purpose
The present study aimed to identify the optimal design of the unicompartmental femoral component through parameter analysis and stability evaluation.
Methods
A finite element (FE) analysis was applied to analyse and adjust the parameter combinations of the anterior tilt angle of the posterior condyle resection surface, the position of the peg, the length of the peg and the inclination angle of the peg, resulting in 10 different FE models. Setting three knee flexion angles of 8.4° (maximum load state during walking), 40° (maximum load state during stair climbing) and 90° (maximum load state during squatting exercise), quantitatively analysing the micromotion values of the bone–prosthesis interface and defining a weighted scoring formula to evaluate the stability of different FE models. The validity of the FE analysis was verified using the Digital Image Correlation (DIC) device.
Results
The errors between the FE analysis and the DIC test at three flexion angles were 5.6%, 1.7% and 11.1%. The 10 different femoral component design models were measured separately. The FE analysis demonstrated that the design with a 0° anterior tilt angle of the posterior condyle resection surface, both pegs placed on the distal resection surface, lengthened 5 mm pegs and a 10° peg inclination angle provided the best stability.
Conclusion
The current study proposed a method for evaluating the stability of the femoral component design. The optimal intersurface stability design of the unicompartmental femoral component was achieved with two pegs placed on the distal resection surface, a 5‐mm peg length increment and a 10° peg inclination. These results might provide a reference for the selection of unicompartmental femoral components in clinical practice and therefore improve the survival rate of future unicompartmental knee arthroplasty. Level of Evidence Level III.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Biomedical Engineering & Med-X Research Institute, Shanghai Jiao Tong University, Shanghai, China.
- Engineering Research Center for Digital Medicine of the Ministry of Education, Shanghai Jiao Tong University, Shanghai, China, Shanghai, China.
- MicroPort Orthopedics, Suzhou, China.
- Department of Sport Medicine, Senior Department of Orthopedics, the Fourth Medical Center of PLA General Hospital, Beijing, China.
- Department of Orthopedics Balgrist University Hospital, Zürich, Switzerland.
- Department of Orthopedic Surgery, Shanghai Sixth People's Hospital Affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai, China.
ORCID-Profile: Zhongyao Li, Rongshan Cheng, Tsung-Yuan Tsai
Förderung
- National Natural Science Foundation of China
- Fundamental Research Funds for the Central Universities
- Science and Technology Innovation Action Plan' of Science and Technology Commission of Shanghai Municipality
- Pudong New Area Science & Technology and Development Fund
Zitieren
APA
Sun, Y., Li, Z., Zhao, K., Dimitriou, D., Yang, B., Cao, Z., Ma, X., Cheng, R., Li, P., & Tsai, T. Y. (2024). Optimal intersurface stability for unicompartmental femoral component design with two pegs placed on the distal resection surface: 5 mm peg length increment and 10° peg inclination. Knee Surgery, Sports Traumatology, Arthroscopy. https://doi.org/10.1002/ksa.12207
Vancouver
Sun Y, Li Z, Zhao K, Dimitriou D, Yang B, Cao Z, et al. Optimal intersurface stability for unicompartmental femoral component design with two pegs placed on the distal resection surface: 5 mm peg length increment and 10° peg inclination. Knee Surgery, Sports Traumatology, Arthroscopy. 2024. doi:10.1002/ksa.12207
BibTeX
@article{Sun2024,
author = {Sun, Yandong and Li, Zhongyao and Zhao, Kaiyu and Dimitriou, Dimitris and Yang, Biao and Cao, Zhangyi and Ma, Xin and Cheng, Rongshan and Li, Pingyue and Tsai, Tsung-Yuan},
title = {Optimal intersurface stability for unicompartmental femoral component design with two pegs placed on the distal resection surface: 5 mm peg length increment and 10° peg inclination},
journal = {Knee Surgery, Sports Traumatology, Arthroscopy},
year = {2024},
doi = {10.1002/ksa.12207}
}
EUVIMED