EUVIMEDEUVIMED entdecken
Crossref

Achieving Specified Laxity in a Noncruciate Total Knee: A Laboratory Design Study

Peter S Walker, Daniel Hennessy, John Perez, Fatema Rahman, Gabriela Zapata, Joseph Bosco

The Journal of Arthroplasty · 2024-08 · 39 (8) · S340-S346

DOI: 10.1016/j.arth.2024.03.015 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2024-08. Verfasst von Peter S Walker und 5 weiteren Autor:innen. Beteiligte Einrichtung: New York University Langone Orthopedic Hospital, New York, New York. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Knee Joint, Joint Instability, Knee Prosthesis. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 32 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Noncruciate total knee arthroplasty designs, including ultracongruent, medially congruent, and medial pivot, are gaining increasing attention in total knee arthroplasty surgery. However, there is no consensus for the bearing surface design, whether there should be different medial, lateral, anterior, and posterior laxities, or whether the medial side should be a medial pivot. This study proposes the criterion of reproducing the laxity of the anatomic knee, defined as the displacements and rotations of the femur on the tibia in the loaded knee when shear and torque are applied. The purpose of this study was to determine the ideal tibial radii to achieve that goal. METHODS: The femoral component was based on the average knee from 100 mild arthritic knee scans. There were 8 tibial components that were designed with different sagittal radii: antero-medial, antero-lateral, postero-medial, and postero-lateral. Radii were defined as the percent height reduction from full conformity with the femoral profile. Components were 3-dimensional-printed. A test rig was constructed where the tibial component was fixed and shear and torque were applied to the femoral component. Displacements and rotations of the femoral component were measured at 0 and 45° of flexion, the latter representing any flexion angle due to the constant femoral sagittal radius. RESULTS: Displacements ranged from 0 to 11 mm, and rotations ranged from 1 to 11°. Anterior femoral displacements were higher than posterior due to the shallow distal-anterior femoral profile. The final femoral and tibial components with the most closely matched anatomic laxity values were designed and tested. CONCLUSIONS: A steeper distal-anterior femoral radius was an advantage. High medial-anterior tibial conformity was important. However, on the lateral side, the posterior sagittal tibial radius had to be shallower than ideal to allow femoral rollback in high flexion. This meant that the posterior laxity displacements on the lateral side were higher than anatomic, and there was no guidance for lateral femoral rollback.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeKnee JointJoint InstabilityKnee ProsthesisProsthesis DesignHumansFemurTibiaRange of Motion, ArticularBiomechanical PhenomenaTorque

Stichwörter der Autor:innen

knee constraintknee laxityknee stabilitymedially congruent kneetotal knee design

Autor:innen und Institutionen

  • New York University Langone Orthopedic Hospital, New York, New York.

Zitieren

APA

Walker, P. S., Hennessy, D., Perez, J., Rahman, F., Zapata, G., & Bosco, J. (2024). Achieving Specified Laxity in a Noncruciate Total Knee: A Laboratory Design Study. The Journal of Arthroplasty, 39(8), S340-S346. https://doi.org/10.1016/j.arth.2024.03.015

Vancouver

Walker PS, Hennessy D, Perez J, Rahman F, Zapata G, Bosco J. Achieving Specified Laxity in a Noncruciate Total Knee: A Laboratory Design Study. The Journal of Arthroplasty. 2024;39(8):S340-S346. doi:10.1016/j.arth.2024.03.015

BibTeX
@article{Walker2024,
  author = {Walker, Peter S and Hennessy, Daniel and Perez, John and Rahman, Fatema and Zapata, Gabriela and Bosco, Joseph},
  title = {Achieving Specified Laxity in a Noncruciate Total Knee: A Laboratory Design Study},
  journal = {The Journal of Arthroplasty},
  year = {2024},
  volume = {39},
  number = {8},
  pages = {S340--S346},
  doi = {10.1016/j.arth.2024.03.015}
}