Handheld ultrasound for portal vein evaluation in pediatric patients

Ulrich Kaiser, Fabian Kück, Christian Stroszczynski, Ernst Michael Jung, Barbara Greiner

Clinical Hemorheology and Microcirculation · 07.11.2024 · 88 (s1) · S57-S68

DOI: 10.3233/ch-248105 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Hemorheology and Microcirculation, erschienen am 07.11.2024. Verfasst von Ulrich Kaiser und 4 weiteren Autor:innen. Beteiligte Einrichtung: Medical Clinic and Polyclinic III, University Hospital Regensburg, Regensburg, Germany. Hauptthemen laut MEDLINE: Portal Vein. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: To assess diagnostic capabilities of a modern handheld ultrasound device (HUD) for portal vein (PV) evaluation in pediatric patients and to verify if age, body mass index (BMI) and object depth correlate to imagequality. METHODS: 45 patients (3.9 months– 17.9 years; 10.3 years±5.4) were examined using a HUD and cart-based high-end ultrasound system (HEUS). The intra-and extrahepatic PV was scanned using B-Mode and Color-coded Doppler sonography (CCDS). A five-point Likert scale was applied to evaluate image quality. Results were interpreted by two readers in consensus. Scores were compared between the scanners and correlated to age, BMI and skin-to-portal-vein-distance (SPVD). The influence on image quality and the difference between the two devices were analyzed using ordinal and Bayesian logistic regression models. RESULTS: ≥4 points (mild or no limitations) were achieved by the HUD in 60% for B-Mode and 56% for CCDS. There was a statistically significant negative correlation between the image quality achieved by the HUD and the SPVD for B-Mode and CCDS. The effect of BMI and SPVD on having≥4 points differed significantly between the HUD and HEUS for B-Mode and for CCDS. CONCLUSIONS: Sufficient to excellent image quality for portal vein assessment in pediatric patients was achieved by the HUD except for the most deep-lying structures. Compared to HEUS, diagnostic performance of the HUD is lower, based on a negative correlation with object depth.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Portal VeinHumansChildChild, PreschoolInfantAdolescentMaleFemaleUltrasonographyBody Mass IndexUltrasonography, Doppler, Color

Stichwörter der Autor:innen

Handheld ultrasoundVscan Airpediatricportal veinvascular

Autor:innen und Institutionen

  • Medical Clinic and Polyclinic III, University Hospital Regensburg, Regensburg, Germany.
  • Department of Medical Statistics, University Medical Center Göttingen, Göttingen, Germany.
  • Institute for Diagnostic Radiology and Interdisciplinary Ultrasound, University Hospital Regensburg, Regensburg, Germany.

Zitieren

APA

Kaiser, U., Kück, F., Stroszczynski, C., Jung, E. M., & Greiner, B. (2024). Handheld ultrasound for portal vein evaluation in pediatric patients. Clinical Hemorheology and Microcirculation, 88(s1), S57-S68. https://doi.org/10.3233/ch-248105

Vancouver

Kaiser U, Kück F, Stroszczynski C, Jung EM, Greiner B. Handheld ultrasound for portal vein evaluation in pediatric patients. Clinical Hemorheology and Microcirculation. 2024;88(s1):S57-S68. doi:10.3233/ch-248105

BibTeX
@article{Kaiser2024,
  author = {Kaiser, Ulrich and Kück, Fabian and Stroszczynski, Christian and Jung, Ernst Michael and Greiner, Barbara},
  title = {Handheld ultrasound for portal vein evaluation in pediatric patients},
  journal = {Clinical Hemorheology and Microcirculation},
  year = {2024},
  volume = {88},
  number = {s1},
  pages = {S57--S68},
  doi = {10.3233/ch-248105}
}
Kostenlos starten →