Pulmonary Vein Anatomy, Cooling Kinetics, and Recurrence After Cryoballoon Ablation for Atrial Fibrillation
Journal of Arrhythmia · 2026-12 · 42 (6)
DOI: 10.1002/joa3.70511 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Arrhythmia, online erschienen am 02.11.2026. Verfasst von Kuo-Ming Yang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Medical Center Kaohsiung Veterans General Hospital Kaohsiung Taiwan. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Cryoballoon pulmonary vein isolation depends on balloon‐to‐ostium contact. We assessed whether computed tomography‐derived pulmonary vein ostial area (OA) and ovality index (OI) were associated with cooling after accounting for pulmonary vein (PV) branch and within‐patient clustering, and whether PV anatomy was associated with recurrence.
Methods
We retrospectively studied 120 patients with symptomatic atrial fibrillation undergoing second‐generation cryoballoon ablation. Vein‐level associations with nadir temperature, 1 min temperature, and time to isolation were assessed using linear mixed‐effects models with a patient random intercept and PV branch as a fixed effect; generalized estimating equations served as sensitivity analyses. Given 23 post‐blanking recurrences, logistic models were limited to left atrial volume index (LAVI) and one clinically selected covariate, with Firth‐penalized regression as a sensitivity analysis.
Results
All patients achieved acute PV isolation. Larger OA was associated with lower nadir temperature ( β = −0.583°C per 50 mm 2 ; 95% CI, −0.910 to −0.257; p < 0.001) and 1 min temperature ( β = −0.441°C per 50 mm 2 ; 95% CI, −0.764 to −0.118; p = 0.008) after branch and clustering adjustment. OI was not associated with either temperature outcome. Twenty‐three patients (19.2%) developed recurrent atrial arrhythmia. LAVI was associated with recurrence (adjusted OR, 1.659 per 10 mL/m 2 ; 95% CI, 1.250–2.201; p < 0.001), whereas persistent AF was not.
Conclusions
OA showed a modest association with cryoballoon temperatures but should not be used as a stand‐alone prognostic marker. LAVI remained associated with recurrence in parsimonious patient‐level analyses.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Medical Center Kaohsiung Veterans General Hospital Kaohsiung Taiwan.
- Department of Critical Care Medicine Kaohsiung Veterans General Hospital Kaohsiung Taiwan.
- Department of Optometry Shu-Zen Junior College of Medicine and Management Kaohsiung Taiwan.
- School of Medicine, College of Medicine National Yang Ming Chiao Tung University Taipei Taiwan.
- Institute of Science Education and Environmental Education National Kaohsiung Normal University Kaohsiung Taiwan.
- Department of Physical Therapy Shu-Zen Junior College of Medicine and Management Kaohsiung Taiwan.
ORCID-Profile: Kuo-Ming Yang, Cheng-Hung Chiang
Zitieren
APA
Yang, K. M., Hsiao, H. C., Yeh, T. C., Chiou, C. W., Hung, J. F., & Chiang, C. H. (2026). Pulmonary Vein Anatomy, Cooling Kinetics, and Recurrence After Cryoballoon Ablation for Atrial Fibrillation. Journal of Arrhythmia, 42(6). https://doi.org/10.1002/joa3.70511
Vancouver
Yang KM, Hsiao HC, Yeh TC, Chiou CW, Hung JF, Chiang CH. Pulmonary Vein Anatomy, Cooling Kinetics, and Recurrence After Cryoballoon Ablation for Atrial Fibrillation. Journal of Arrhythmia. 2026;42(6). doi:10.1002/joa3.70511
BibTeX
@article{Yang2026,
author = {Yang, Kuo-Ming and Hsiao, Hsiang-Chiang and Yeh, Tung-Chen and Chiou, Chun-Wang and Hung, Jeng-Fung and Chiang, Cheng-Hung},
title = {Pulmonary Vein Anatomy, Cooling Kinetics, and Recurrence After Cryoballoon Ablation for Atrial Fibrillation},
journal = {Journal of Arrhythmia},
year = {2026},
volume = {42},
number = {6},
doi = {10.1002/joa3.70511}
}
EUVIMED