Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein Gaps

Ryuta Watanabe iD, Koichi Nagashima iD, Hikaru Masuda, Akinori Matsushima, Shu Hirata iD, Yuji Saito iD, Masanaru Sawada, Yasuo Okumura iD

Journal of Arrhythmia · 2026-04 · 42 (2)

DOI: 10.1002/joa3.70333 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Arrhythmia, online erschienen am 04.04.2026. Verfasst von Ryuta Watanabe und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Department of Medicine Nihon University School of Medicine Tokyo Japan. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Pulsed field ablation (PFA) for atrial fibrillation (AF) requires adequate catheter–tissue contact. The Contact Index visualizes impedance change for the pentaspline catheter, but a validated contact threshold is unknown.

Objective

The objective of this study is to define an impedance‐change cutoff that predicts sufficient contact to prevent acute pulmonary vein gaps (PVG) during pulmonary vein isolation (PVI), and to validate Contact–Index‐guided PVI.

Methods

A prospective, single‐center, two‐phase study was performed. Phase 1 ( n = 10) measured impedance change (%) at eight antral segments per vein immediately before PFA in basket and flower configurations; the larger per‐segment value was analyzed. Post‐PVI high‐density voltage mapping (PVG ≥ 0.5 mV) was the reference; a receiver operating characteristic analysis identified the cutoff. Phase 2 ( n = 26) delivered energy only when all target splines met/exceeded the cutoff value obtained from Phase 1.

Results

In Phase 1, PVG occurred in 13/320 segments across 2/10 patients (4/40 veins). Impedance change was lower at PVG‐positive than PVG‐negative segments (median 2 [0–6.5]% vs. 24 [17–32]%, p < 0.001). The cutoff value obtained from Phase 1 was ≥ 10% (area under the curve 0.99; sensitivity 100%; specificity 92.1%). In Phase 2, enforcing ≥ 10% yielded no patient‐level PVG (0/26 vs. 2/10; p = 0.07) and fewer gaps per vein (0/104 vs. 4/40; p = 0.005) and per segment (0/832 vs. 13/320; p < 0.001) compared with Phase 1. Procedure time, PVI time, fluoroscopy time, and total applications were similar between phases.

Conclusions

An impedance‐change threshold of ≥ 10% on the Contact Index reliably indicates adequate contact with the pentaspline catheter and, when enforced, reduces acute PVG without prolonging procedures.

Schlagwörter

Stichwörter der Autor:innen

Contact Indexatrial fibrillationimpedancepentaspline catheterpulmonary vein gapspulsed field ablation

Autor:innen und Institutionen

  • Division of Cardiology, Department of Medicine Nihon University School of Medicine Tokyo Japan.

ORCID-Profile: Ryuta Watanabe, Koichi Nagashima, Shu Hirata, Yuji Saito, Yasuo Okumura

Zitieren

APA

Watanabe, R., Nagashima, K., Masuda, H., Matsushima, A., Hirata, S., Saito, Y., Sawada, M., & Okumura, Y. (2026). Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein Gaps. Journal of Arrhythmia, 42(2). https://doi.org/10.1002/joa3.70333

Vancouver

Watanabe R, Nagashima K, Masuda H, Matsushima A, Hirata S, Saito Y, et al. Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein Gaps. Journal of Arrhythmia. 2026;42(2). doi:10.1002/joa3.70333

BibTeX
@article{Watanabe2026,
  author = {Watanabe, Ryuta and Nagashima, Koichi and Masuda, Hikaru and Matsushima, Akinori and Hirata, Shu and Saito, Yuji and Sawada, Masanaru and Okumura, Yasuo},
  title = {Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein Gaps},
  journal = {Journal of Arrhythmia},
  year = {2026},
  volume = {42},
  number = {2},
  doi = {10.1002/joa3.70333}
}
Kostenlos starten →