Development and Initial Evaluation of an In‐House Paediatric CT Immobilisation Device

Maeve Masterson iD, Vinay Krishnan, Ahilan Kuganesan, Mohamed K Badawy iD

Journal of Medical Radiation Sciences · 2025

DOI: 10.1002/jmrs.891 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Medical Radiation Sciences, 2025. Verfasst von Maeve Masterson und 3 weiteren Autor:innen. Beteiligte Einrichtung: Monash Health Imaging, Monash Health, Clayton, Victoria, Australia. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Immobilization. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Paediatric imaging often requires immobilisation to reduce motion artefacts, reduce the risk of repeat imaging and subsequently reduce patient radiation exposure. Commercial immobilisation devices may be expensive and infrequently used, making them difficult to justify for institutions with budget constraints. This study aimed to design and evaluate a cost‐effective, in‐house alternative.

Methods

An in‐house CT immobilisation device was developed using acrylic and polyethylene foam, tailored for newborns under 5 kg. Image quality was assessed using a Gammex ACR phantom on a CT scanner. Hounsfield units, noise and artefacts were compared between the in‐house and commercial devices. Artefacts were independently evaluated by a radiographer and a medical physicist.

Results

The in‐house device performed comparably to the commercial version, with uniformity deviation within 2 HU and noise within 2.4 SD of baseline values; artefacts were minimal for both devices. There was no significant difference between the two immobilisation devices ( p = 0.711). Stability concerns with the in‐house device's square base could be resolved using immobilisation straps, and future iterations should address the foam insert's porous nature.

Conclusion

The in‐house CT immobilisation device offers a cost‐effective alternative without compromising image quality, making it a viable option for healthcare institutions with limited budgets. However, its implementation would require further clinical evaluation and compliance with local regulations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedImmobilizationHumansPhantoms, ImagingArtifactsEquipment DesignInfant, Newborn

Stichwörter der Autor:innen

computed tomographyimage qualityimmobilisationpaediatricradiation protection

Autor:innen und Institutionen

  • Monash Health Imaging, Monash Health, Clayton, Victoria, Australia.
  • Department of Medical Imaging and Radiation Sciences, School of Primary and Allied Health Care, Faculty of Medicine, Nursing and Health Sciences, Monash University, Clayton, Victoria, Australia.

ORCID-Profile: Maeve Masterson, Mohamed K Badawy

Zitieren

APA

Masterson, M., Krishnan, V., Kuganesan, A., & Badawy, M. K. (2025). Development and Initial Evaluation of an In‐House Paediatric CT Immobilisation Device. Journal of Medical Radiation Sciences. https://doi.org/10.1002/jmrs.891

Vancouver

Masterson M, Krishnan V, Kuganesan A, Badawy MK. Development and Initial Evaluation of an In‐House Paediatric CT Immobilisation Device. Journal of Medical Radiation Sciences. 2025. doi:10.1002/jmrs.891

BibTeX
@article{Masterson2025,
  author = {Masterson, Maeve and Krishnan, Vinay and Kuganesan, Ahilan and Badawy, Mohamed K},
  title = {Development and Initial Evaluation of an In‐House Paediatric CT Immobilisation Device},
  journal = {Journal of Medical Radiation Sciences},
  year = {2025},
  doi = {10.1002/jmrs.891}
}
Kostenlos starten →