Iodine concentration, HU accuracy, and metal artifacts evaluation on second-generation dual-layer spectral detector CT images with metal implants: a phantom study
Acta Radiologica · 2024
DOI: 10.1177/02841851241269853 ↗Kurz gefasst
Fachartikel in der Zeitschrift Acta Radiologica, 2024. Verfasst von Xinming Zhang und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Engineering, Nanfang Hospital, Southern Medical University, Guangzhou, PR China. Hauptthemen laut MEDLINE: Phantoms, Imaging, Artifacts, Metals, Tomography, X-Ray Computed. Die Arbeit nennt 28 Literaturangaben.
Abstract (Originalsprache)
Background
Metal implants may affect the image quality, iodine concentration (IC), and CT Hounsfield unit (HU) quantification accuracy.
Purpose
To investigate the quantitative accuracy of IC and HU from dual-layer spectral detector (DLCT) in the presence of metal artifacts. Material and Methods An experimental cylindrical phantom containing eight iodine inserts and two metal inserts was designed. The phantom underwent scanning at three radiation dose levels and two tube voltage settings. A set of conventional images (CIs), virtual monoenergetic images (VMIs), and iodine concentration maps (ICMs) were generated and measured for all the eight iodine inserts. Quantitative indicators of mean absolute percentage error (MAPE), artifact index (AI), contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and standard deviation (SD) on CIs and VMIs were calculated for IC and HU. Subjective score evaluation was also conducted.
Results
The MAPE iodine values of all regions of interest across different scanning configurations were all
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Clinical Engineering, Nanfang Hospital, Southern Medical University, Guangzhou, PR China.
- School of Biomedical Engineering, Southern Medical University, Guangzhou, PR China.
- Clinical and Technical Support, Philips Healthcare, Shanghai, PR China.
ORCID-Profile: Xinming Zhang
Zitieren
APA
Zhang, X., Li, H., Wang, H., Liu, X., Deng, W., Zhong, J., Fan, L., Ling, Q., Hu, D., Qi, H., & Chen, H. (2024). Iodine concentration, HU accuracy, and metal artifacts evaluation on second-generation dual-layer spectral detector CT images with metal implants: a phantom study. Acta Radiologica. https://doi.org/10.1177/02841851241269853
Vancouver
Zhang X, Li H, Wang H, Liu X, Deng W, Zhong J, et al. Iodine concentration, HU accuracy, and metal artifacts evaluation on second-generation dual-layer spectral detector CT images with metal implants: a phantom study. Acta Radiologica. 2024. doi:10.1177/02841851241269853
BibTeX
@article{Zhang2024,
author = {Zhang, Xinming and Li, Hanwei and Wang, Haowen and Liu, Xiaomin and Deng, Weiwei and Zhong, Jianqiu and Fan, Litian and Ling, Qingqing and Hu, Debin and Qi, Hongliang and Chen, Hongwen},
title = {Iodine concentration, HU accuracy, and metal artifacts evaluation on second-generation dual-layer spectral detector CT images with metal implants: a phantom study},
journal = {Acta Radiologica},
year = {2024},
doi = {10.1177/02841851241269853}
}
EUVIMED