Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2026. Verfasst von Jeanette C Deck und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Balgrist University Hospital, Zurich, Switzerland. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Artifacts, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Slice encoding for metal artifact correction (SEMAC) is used in MRI for total hip arthroplasties for metal artifact reduction. However, ripple artifacts often arise near the implant, impairing diagnosis in affected regions. Hypothesis Applying a wavelet domain filter (WD‐Filter) might reduce ripple artifacts and improve image quality. Study Type Retrospective. Subjects 100 patients with primary total hip arthroplasty (female: 51, mean age: (64 ± 13) years; male: 49, mean age: (65 ± 13) years) who underwent clinically indicated MRI using compressed sensing SEMAC. Field Strength/Sequences 1.5 T; compressed sensing‐accelerated turbo spin‐echo (TSE)‐based SEMAC sequence. Assessment Three readers rated 250 ripple artifacts before and after applying the WD‐filter. 4‐point Likert scales were used for the intensity of ripple artifacts (Likert‐RA, “none”: 1, “mild”: 2, “moderate”: 3, “severe”: 4) and the image quality (Likert‐IQ, “poor”: 1, “moderate”: 2, “good”: 3, “excellent”: 4). The relative reduction of the signal intensity standard deviation () was measured in a region of interest (ROI) containing the ripple artifact. Statistical Tests Inter‐reader agreement was assessed with Fleiss' kappa, with 95% confidence intervals. Differences between pre‐ and post‐filter ratings (Likert‐RA and Likert‐IQ) were evaluated using the Wilcoxon signed‐rank test, while was compared using a two‐sided paired t ‐test. A p value of less than 0.05 indicated statistical significance.
Results
Prior to WD‐Filter application, ripple artifacts were rated as “severe”, as “moderate”, and as “mild”. After filtering, artifacts were rated as “none”, and as “mild”. Upon filter application, Likert‐RA and Likert‐IQ significantly improved from to and to , respectively, with a large effect size ( r > 0.89) and “almost perfect” inter‐rater agreement (kappa range: 0.86–0.96). The filtered images showed a significantly reduced compared to the unfiltered images (relative change: ). Data Conclusion The WD‐Filter effectively reduced ripple artifacts in MRI of total hip arthroplasty and improved image quality. Evidence Level 3. Technical Efficacy Stage 3.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Balgrist University Hospital, Zurich, Switzerland.
- Medical Faculty, University of Zurich (UZH), Zurich, Switzerland.
- Swiss Innovation Hub (SIH), Siemens Healthineers International AG, Zurich, Switzerland.
- Department of Orthopedic Surgery, Balgrist University Hospital, Zurich, Switzerland.
- Swiss Center for Musculoskeletal Imaging (SCMI), Balgrist Campus AG, Zurich, Switzerland.
ORCID-Profile: Jeanette C Deck, Sophia S Goller, Georg W Kajdi, Adrian A Marth, Constantin von Deuster, Reto Sutter
Zitieren
APA
Deck, J. C., Goller, S. S., Kajdi, G. W., Marth, A. A., Zingg, P., von Deuster, C., & Sutter, R. (2026). MRI of Total Hip Arthroplasty: Successful Ripple Artifact Reduction Using Wavelet Domain Filtering. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.70448
Vancouver
Deck JC, Goller SS, Kajdi GW, Marth AA, Zingg P, von Deuster C, et al. MRI of Total Hip Arthroplasty: Successful Ripple Artifact Reduction Using Wavelet Domain Filtering. Journal of Magnetic Resonance Imaging. 2026. doi:10.1002/jmri.70448
BibTeX
@article{Deck2026,
author = {Deck, Jeanette C and Goller, Sophia S and Kajdi, Georg W and Marth, Adrian A and Zingg, Patrick and von Deuster, Constantin and Sutter, Reto},
title = {MRI of Total Hip Arthroplasty: Successful Ripple Artifact Reduction Using Wavelet Domain Filtering},
journal = {Journal of Magnetic Resonance Imaging},
year = {2026},
doi = {10.1002/jmri.70448}
}
EUVIMED