EUVIMEDEUVIMED entdecken
Crossref

The Minimum Admissible Detuning Efficiency of MRI Receive‐Only Surface Coils

Sina Marhabaie iD, Aimé Labbé, Bruno Quesson, Marie Poirier-Quinot

Journal of Magnetic Resonance Imaging · 2024

DOI: 10.1002/jmri.29209 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2024. Verfasst von Sina Marhabaie und 3 weiteren Autor:innen. Beteiligte Einrichtung: Université Paris-Saclay, CNRS, Inserm, Laboratoire d'Imagerie Biomédicale Multimodale Paris Saclay, Orsay, France. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Equipment Design, Computer Simulation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Gefördert von: French National Research Agency, France Life Imaging, Agence Nationale de la Recherche.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The minimum admissible detuning efficiency (DE) of a receive coil is an essential parameter for coil designers. A receive coil with inefficient detuning leads to inhomogeneous B 1 during excitation. Previously proposed criteria for quantifying the DE rely on indirect measurements and are difficult to implement.

Purpose

To present an alternative method to quantify the DE of receive‐only surface coils. Study Type Theoretical study supported by simulations and phantom experiments. Phantoms Uniform spherical (100 mm diameter) and cylindrical (66 mm diameter) phantoms. Field Strength/Sequence Dual repetition time B 1 mapping sequence at 1.5T, and Bloch‐Siegert shift B 1 mapping sequence at 3.0T. Assessment One non‐planar (80 × 43 mm 2 ) and two planar (40 and 57 mm diameter) surface coils were built. Theoretical analysis was performed to determine the minimum DE required to avoid B 1 distortions. Experimental B 1 maps were acquired for the non‐planar and planar surface coils at both 1.5T and 3.0T and visually compared with simulated B 1 maps to assess the validity of the theoretical analysis. Statistical Tests None.

Results

Based on the theoretical analysis, the proposed minimum admissible DE, defined as DE thr = 20 Log ( Q ) + 13 dB, depended only on the quality factor ( Q ) of the coil and was independent of coil area and field strength. Simulations and phantom experiments showed that when the DE was higher than this minimum threshold level, the B 1 field generated by the transmission coil was not modified by the receive coil. Data Conclusion The proposed criterion for assessing the DE is simple to measure, and does not depend on the area of the coil or on the magnetic field strength, up to 3T. Experimental and simulated B 1 maps confirmed that detuning efficiencies above the theoretically derived minimal admissible DE resulted in a non‐distorted B 1 field. Evidence Level 2 Technical Efficacy Stage 1

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMagnetic Resonance ImagingEquipment DesignComputer SimulationReproducibility of ResultsHumansImage Processing, Computer-AssistedAlgorithmsModels, Theoretical

Stichwörter der Autor:innen

B1 distortionB1 inhomogeneityQ factorcoil decouplingdetuningdetuning efficiencyquality factorsurface coil

Autor:innen und Institutionen

  • Université Paris-Saclay, CNRS, Inserm, Laboratoire d'Imagerie Biomédicale Multimodale Paris Saclay, Orsay, France.
  • University of Bordeaux, CNRS, CRMSB, UMR 5536, IHU Liryc, Bordeaux, France.

ORCID-Profile: Sina Marhabaie

Förderung

  • French National Research Agency
  • France Life Imaging
  • Agence Nationale de la Recherche

Zitieren

APA

Marhabaie, S., Labbé, A., Quesson, B., & Poirier-Quinot, M. (2024). The Minimum Admissible Detuning Efficiency of MRI Receive‐Only Surface Coils. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.29209

Vancouver

Marhabaie S, Labbé A, Quesson B, Poirier-Quinot M. The Minimum Admissible Detuning Efficiency of MRI Receive‐Only Surface Coils. Journal of Magnetic Resonance Imaging. 2024. doi:10.1002/jmri.29209

BibTeX
@article{Marhabaie2024,
  author = {Marhabaie, Sina and Labbé, Aimé and Quesson, Bruno and Poirier-Quinot, Marie},
  title = {The Minimum Admissible Detuning Efficiency of MRI Receive‐Only Surface Coils},
  journal = {Journal of Magnetic Resonance Imaging},
  year = {2024},
  doi = {10.1002/jmri.29209}
}
Kostenlos starten →