Impact of Different Pre‐Dilution Methods on Routine Blood Test Results: A Study With Sysmex XN ‐1000

Lingsong Wu, Mengzhe Qin, Zhe Xing, Fengli Jiang, Sien Zhan, Jintian Hu iD

Journal of Clinical Laboratory Analysis · 2026-07 · 40 (14)

DOI: 10.1002/jcla.70260 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Laboratory Analysis, online erschienen am 02.06.2026. Verfasst von Lingsong Wu und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Laboratory, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China. Hauptthemen laut MEDLINE: Hematologic Tests. Die Arbeit nennt 9 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To compare three manual pre‐dilution procedures with the automated undiluted mode on the Sysmex XN‐1000 analyzer and to identify an optimal standard operating procedure for pre‐dilution testing.

Methods

A total of 60 residual EDTA‐K3 whole‐blood specimens were consecutively collected between March and May 2024 and analyzed within 2 h. Each specimen was tested in the automated undiluted mode and by three manual pre‐dilution procedures: direct mixing, dry‐tip wiping, and wet‐tip wiping. Pre‐dilution was performed at 1:7 (20 μL of blood plus 120 μL of Sysmex DCL diluent) using calibrated pipettes. All procedures were performed by one trained technician. Each method was tested in triplicate, and mean values were analyzed. White blood cell count (WBC), red blood cell count (RBC), hemoglobin (HGB), platelet count (PLT), and hematocrit (HCT) were evaluated by paired t ‐test, Pearson correlation, Passing–Bablok regression, and bias analysis against CLIA allowable total error (TEa) criteria.

Results

For WBC, RBC, HGB, and PLT, no significant difference was found between any pre‐dilution method and the undiluted mode (all p > 0.05); for HCT, only Method 1 differed significantly ( p = 0.022). Passing–Bablok regression showed that the wet‐tip wiping method had the best overall agreement with the undiluted mode. TEa‐based analysis showed that only Methods 2 and 3 met allowable bias limits for all five parameters.

Conclusion

The wet‐tip wiping method showed the best overall analytical performance and may be preferred for low‐volume specimens on the Sysmex XN‐1000 platform.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hematologic TestsHumansBlood Cell CountHematocritReproducibility of ResultsHemoglobinsPlatelet Count

Stichwörter der Autor:innen

Bland–Altman analysisSysmex XN‐1000blood routine testpre‐dilution methodswet‐tip wiping

Substanzen

Hemoglobins

Autor:innen und Institutionen

  • Department of Medical Laboratory, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Department of Cosmetic Injection Center, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.

ORCID-Profile: Jintian Hu

Zitieren

APA

Wu, L., Qin, M., Xing, Z., Jiang, F., Zhan, S., & Hu, J. (2026). Impact of Different Pre‐Dilution Methods on Routine Blood Test Results: A Study With Sysmex XN ‐1000. Journal of Clinical Laboratory Analysis, 40(14). https://doi.org/10.1002/jcla.70260

Vancouver

Wu L, Qin M, Xing Z, Jiang F, Zhan S, Hu J. Impact of Different Pre‐Dilution Methods on Routine Blood Test Results: A Study With Sysmex XN ‐1000. Journal of Clinical Laboratory Analysis. 2026;40(14). doi:10.1002/jcla.70260

BibTeX
@article{Wu2026,
  author = {Wu, Lingsong and Qin, Mengzhe and Xing, Zhe and Jiang, Fengli and Zhan, Sien and Hu, Jintian},
  title = {Impact of Different Pre‐Dilution Methods on Routine Blood Test Results: A Study With Sysmex XN ‐1000},
  journal = {Journal of Clinical Laboratory Analysis},
  year = {2026},
  volume = {40},
  number = {14},
  doi = {10.1002/jcla.70260}
}
Kostenlos starten →