Overcoming Platelet Aggregation Interference: Evaluation of the Complete Blood Count, Differential, and Reticulocyte (CDR) Mode on an Automated Analyzer for Accurate Platelet Counting
The Journal of Applied Laboratory Medicine · 04.11.2026 · 11 (6) · 1432-1441
DOI: 10.1093/jalm/jfag092 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 22.07.2026. Verfasst von Jia-le Gong und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China. Hauptthemen laut MEDLINE: Platelet Aggregation, Blood Platelets. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
This study assessed the ability of the complete blood count, differential, and reticulocyte (CDR) mode of the Mindray BC-6800 Plus analyzer to correct platelet aggregation by disaggregating platelets, evaluating how aggregation severity impacts accuracy to define an objective threshold for resampling.
Methods
Agreement was assessed between CDR mode and the electrical impedance method for platelet counts in normal specimens using Pearson correlation, Passing–Bablok regression, and Bland–Altman analysis. The CDR mode’s ability to correct false elevations caused by microcytes/fragments and false reductions caused by large platelets was evaluated by comparing CDR channel platelet count results (PLT-OF) with smear-estimated values (PLT-M) and fluorescence-method results (PLT-F). Disaggregation efficacy in aggregated samples was analyzed using the MC-80 digital morphology system. The optimal disaggregation threshold was determined by ROC analysis.
Results
In normal specimens, PLT-OF showed a significant correlation with the electrical impedance-method platelet count results (PLT-I) (r = 0.999, P < 0.001). Passing–Bablok regression (PLT-I = 1.00 × PLT-OF) and Bland–Altman analysis demonstrated excellent agreement. For samples with PLT-I false abnormalities, the CDR mode effectively corrected both false elevations (due to microcytes/fragments) and false reductions (due to large platelets), showing no significant differences between PLT-OF and PLT-M or PLT-F (P > 0.05). In platelet-aggregated specimens, PLT-OF was significantly higher than PLT-I (P < 0.001). ROC analysis established 162 aggregates as the optimal threshold for effective disaggregation [area under the curve (AUC) = 0.776]. The dissociation rate was significantly greater in the mild vs severe-aggregation group (P = 0.004). Combining aggregate count with a PLT-OF to PLT-I difference (>115 × 109/L) improved prediction accuracy (AUC = 0.841).
Conclusion
The CDR channel of the Mindray BC-6800 Plus analyzer effectively corrects spurious platelet counts.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Laboratory Medicine, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
Zitieren
APA
Gong, J. L., Liao, Q., Duan, C. J., FengHuang, Z., & Zhang, B. (2026). Overcoming Platelet Aggregation Interference: Evaluation of the Complete Blood Count, Differential, and Reticulocyte (CDR) Mode on an Automated Analyzer for Accurate Platelet Counting. The Journal of Applied Laboratory Medicine, 11(6), 1432-1441. https://doi.org/10.1093/jalm/jfag092
Vancouver
Gong JL, Liao Q, Duan CJ, FengHuang Z, Zhang B. Overcoming Platelet Aggregation Interference: Evaluation of the Complete Blood Count, Differential, and Reticulocyte (CDR) Mode on an Automated Analyzer for Accurate Platelet Counting. The Journal of Applied Laboratory Medicine. 2026;11(6):1432-1441. doi:10.1093/jalm/jfag092
BibTeX
@article{Gong2026,
author = {Gong, Jia-le and Liao, Qing and Duan, Chao-Jun and FengHuang, Zhouzhou and Zhang, Bo},
title = {Overcoming Platelet Aggregation Interference: Evaluation of the Complete Blood Count, Differential, and Reticulocyte (CDR) Mode on an Automated Analyzer for Accurate Platelet Counting},
journal = {The Journal of Applied Laboratory Medicine},
year = {2026},
volume = {11},
number = {6},
pages = {1432--1441},
doi = {10.1093/jalm/jfag092}
}
EUVIMED