EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

V‐ PRO Blood Collection Tubes: Validation for Clinical Chemistry and Immunoassay Tests

Anwar Borai iD, Wedyan Alsharif, Amirah Alhindi, Maha Alqahtani, Mohieldin Elsayid, Haitham Khalil, Salwa Al Marwani, Abobaker Yagoot, Janet Magjacot, Maha Al Meteiri, Rawan Alyamani, Hind Abdulhakim und 1 weitere

Journal of Clinical Laboratory Analysis · 2025

DOI: 10.1002/jcla.70007 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Clinical Laboratory Analysis, 2025. Verfasst von Anwar Borai und 12 weiteren Autor:innen. Beteiligte Einrichtung: King Abdullah International Medical Research Center (KAIMRC), King Saud bin Abdulaziz University for Health Sciences (KSAU-HS), King Abdulaziz Medica…. Hauptthemen laut MEDLINE: Blood Specimen Collection, Blood Chemical Analysis. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Becton Dickenson and Company.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In accredited laboratories, each component of diagnostic products—such as laboratory instruments, reagents, and blood collection tubes must be validated before integration into routine patient testing. BD Vacutainers are commonly used in clinical laboratories compared to other blood collection tubes, while V‐PRO tubes have recently been introduced to the market without prior laboratory validation. This study compares V‐PRO tubes to BD Vacutainers to assess the validity of using V‐PRO tubes for blood testing. Materials and Methods Blood samples were collected simultaneously into two different brands of tubes (V‐PRO and BD) from 60 subjects. A standardized procedure was employed for sample collection, and analysis. A total of 28 chemistry tests and 20 immunoassays were analyzed using Abbott instruments, while high‐performance liquid chromatography was used for testing glycated hemoglobin. The biases of V‐PRO compared to BD were evaluated against current desirable quality specifications for bias derived from biological variation. For technical validation, a designated survey was distributed to various institutes using both tube types in their laboratories.

Results

The V‐PRO tube exhibited biases exceeding the desirable limits for CO 2 (3.2%), magnesium (2.0%), thyroid‐stimulating hormone (11.7%), and estradiol (−8.5%). Survey results indicated a higher percentage of major pre‐analytical, analytical, and post‐analytical errors when using the V‐PRO tube compared to the BD Vacutainer.

Conclusions

Laboratories currently using BD vacutainers should exercise caution if they intend to perform chemistry and immunoassay tests with V‐PRO tubes. The technical validation outcomes for V‐PRO were not acceptable due to significant faults identified in comparison to BD Vacutainer.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood Specimen CollectionBlood Chemical AnalysisHumansImmunoassayReproducibility of Results

Stichwörter der Autor:innen

V‐PRObloodtubevacutainervalidation

Autor:innen und Institutionen

  • King Abdullah International Medical Research Center (KAIMRC), King Saud bin Abdulaziz University for Health Sciences (KSAU-HS), King Abdulaziz Medical City, Ministry of National Guard, Jeddah, Saudi Arabia.
  • King Fahad Armed Forces Hospital, Department of Pathology and Laboratory Medicine, Jeddah, Saudi Arabia.

ORCID-Profile: Anwar Borai, Majid Al-Thaqafy

Förderung

  • Becton Dickenson and Company

Zitieren

APA

Borai, A., Alsharif, W., Alhindi, A., Alqahtani, M., Elsayid, M., Khalil, H., Al Marwani, S., Yagoot, A., Magjacot, J., Al Meteiri, M., Alyamani, R., Abdulhakim, H., & Al-Thaqafy, M. (2025). V‐ PRO Blood Collection Tubes: Validation for Clinical Chemistry and Immunoassay Tests. Journal of Clinical Laboratory Analysis. https://doi.org/10.1002/jcla.70007

Vancouver

Borai A, Alsharif W, Alhindi A, Alqahtani M, Elsayid M, Khalil H, et al. V‐ PRO Blood Collection Tubes: Validation for Clinical Chemistry and Immunoassay Tests. Journal of Clinical Laboratory Analysis. 2025. doi:10.1002/jcla.70007

BibTeX
@article{Borai2025,
  author = {Borai, Anwar and Alsharif, Wedyan and Alhindi, Amirah and Alqahtani, Maha and Elsayid, Mohieldin and Khalil, Haitham and Al Marwani, Salwa and Yagoot, Abobaker and Magjacot, Janet and Al Meteiri, Maha and Alyamani, Rawan and Abdulhakim, Hind and Al-Thaqafy, Majid},
  title = {V‐ PRO Blood Collection Tubes: Validation for Clinical Chemistry and Immunoassay Tests},
  journal = {Journal of Clinical Laboratory Analysis},
  year = {2025},
  doi = {10.1002/jcla.70007}
}
Kostenlos starten →