EUVIMEDEUVIMED entdecken
Vergleichsstudie

Platelet Loss and Collection Efficiency During Autologous PBSC Collection: A Retrospective Comparison of Two Apheresis Platforms

Kamolchanok Kulchutisin iD, Thanatphak Warindpong iD, Parichart Permpikul iD, Kulvara Kittisares iD, Chanchira Noothong, Nongnuch Choodokput, Chanok Ngamsanga, Sirikwan Boonmoh, Orathai Taka, Weerapat Owattanapanich iD, Janejira Kittivorapart iD

Journal of Clinical Apheresis · 2026

DOI: 10.1002/jca.70200 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Clinical Apheresis, 2026. Verfasst von Kamolchanok Kulchutisin und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Transfusion Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand. Hauptthemen laut MEDLINE: Blood Component Removal, Blood Platelets, Peripheral Blood Stem Cells, Leukapheresis. Die Arbeit nennt 17 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Hematopoietic stem cell transplantation is a standard treatment for hematologic malignancies, and peripheral blood stem cell (PBSC) collection by leukapheresis is the preferred harvesting method. Platelet loss is a common procedure‐related complication. This study compared platelet loss and product characteristics between two PBSC apheresis systems and evaluated the pre‐collection platelet baseline needed to avoid procedure‐related thrombocytopenia. This retrospective cohort study was conducted at a tertiary care hospital from March 2024 to October 2025. We reviewed 158 autologous PBSC collections in 110 patients using Spectra Optia in continuous mononuclear cell mode and 63 collections in 55 patients using Amicus in Mononuclear Cell mode. Platelet loss, hemoglobin, hematocrit, white blood cell count, granulocyte percentage, and CD34 + product characteristics were evaluated. Platelet loss was lower with Amicus than with Spectra Optia (median [IQR] 19.28% [10.94%–31.43%] vs. 33.33% [24.40%–42.50%]; p < 0.001). Spectra Optia had higher collection efficiency (48.30% [37.03%–63.64%] vs. 37.23% [27.26%–48.58%]; p < 0.001), higher product volume (257 [238–283] vs. 177 [155–202] mL; p < 0.001), and higher platelet concentration (515 [390–680] vs. 340 [240–480] × 10 3 /μL; p < 0.001). A pre‐collection platelet threshold of 35 000/μL was proposed to maintain post‐collection counts ≥ 20 000/μL. Both systems were effective for autologous PBSC collection. Spectra Optia provided higher collection efficiency, whereas Amicus was associated with lower platelet loss. Appropriate pre‐apheresis assessment of CD34 + cells and platelet count may help optimize collection and reduce procedure‐related thrombocytopenia.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood Component RemovalBlood PlateletsPeripheral Blood Stem CellsLeukapheresisHumansRetrospective StudiesFemaleMalePlatelet CountMiddle AgedAdultPeripheral Blood Stem Cell Transplantation

Stichwörter der Autor:innen

AmicusSpectra Optiaapheresiscell separatorhematopoietic stem cell transplantationperipheral blood stem cell collectionplatelet loss

Autor:innen und Institutionen

  • Department of Transfusion Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
  • Center of Excellence of Siriraj Adult Acute Myeloid/Lymphoblastic Leukemia, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
  • Division of Hematology, Department of Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.

ORCID-Profile: Kamolchanok Kulchutisin, Thanatphak Warindpong, Parichart Permpikul, Kulvara Kittisares, Weerapat Owattanapanich, Janejira Kittivorapart

Zitieren

APA

Kulchutisin, K., Warindpong, T., Permpikul, P., Kittisares, K., Noothong, C., Choodokput, N., Ngamsanga, C., Boonmoh, S., Taka, O., Owattanapanich, W., & Kittivorapart, J. (2026). Platelet Loss and Collection Efficiency During Autologous PBSC Collection: A Retrospective Comparison of Two Apheresis Platforms. Journal of Clinical Apheresis. https://doi.org/10.1002/jca.70200

Vancouver

Kulchutisin K, Warindpong T, Permpikul P, Kittisares K, Noothong C, Choodokput N, et al. Platelet Loss and Collection Efficiency During Autologous PBSC Collection: A Retrospective Comparison of Two Apheresis Platforms. Journal of Clinical Apheresis. 2026. doi:10.1002/jca.70200

BibTeX
@article{Kulchutisin2026,
  author = {Kulchutisin, Kamolchanok and Warindpong, Thanatphak and Permpikul, Parichart and Kittisares, Kulvara and Noothong, Chanchira and Choodokput, Nongnuch and Ngamsanga, Chanok and Boonmoh, Sirikwan and Taka, Orathai and Owattanapanich, Weerapat and Kittivorapart, Janejira},
  title = {Platelet Loss and Collection Efficiency During Autologous PBSC Collection: A Retrospective Comparison of Two Apheresis Platforms},
  journal = {Journal of Clinical Apheresis},
  year = {2026},
  doi = {10.1002/jca.70200}
}
Kostenlos starten →