Is Accurate Dental Implant Placement Feasible Using a Novel Dynamic Computer‐Assisted Surgery System Without Patient Optical Markers or Registration? A Preliminary Retrospective Cohort Study
Clinical and Experimental Dental Research · 2026
DOI: 10.1002/cre2.70256 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical and Experimental Dental Research, 2026. Verfasst von Gang He und 4 weiteren Autor:innen. Beteiligte Einrichtung: Sunshine Dental Clinic, Ningbo, China. Hauptthemen laut MEDLINE: Surgery, Computer-Assisted, Dental Implantation, Endosseous, Dental Implants. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Objectives
To evaluate the accuracy and surgery time of dental implant placement using a novel dynamic computer‐assisted implant surgery (dCAIS) system that eliminates the need for patient registration and optical tracking markers. The secondary objective was to compare these outcomes with those obtained using a conventional dCAIS system. Materials and Methods A preliminary retrospective cohort study was conducted involving 33 participants (33 implants). Eleven implants were placed using the novel dCAIS system that determines patient positioning based on anterior tooth anatomy (Prototype group), while 22 implants were placed using a conventional dCAIS system requiring standard registration and an optical tracking marker (Control group). Pre‐ and postoperative cone‐beam computed tomography (CBCT) scans were superimposed to assess implant placement accuracy. Descriptive and bivariate analyses were performed to compare accuracy and surgery time between the two groups.
Results
Mean angular deviations were similar between groups ( p = 0.924): 1.95° (SD 1.38) in the prototype group and 2.38° (SD 2.30) in the control group. No significant differences were observed in platform global deviation (mean difference [MD]: −0.33 mm; 95% CI: −0.75 to 0.09), apex global deviation (MD: −0.43 mm; 95% CI: −0.94 to 0.08), or apex depth deviation (MD: 0.28 mm; 95% CI: −0.30 to 0.86). Surgical procedures were significantly faster in the prototype group ( p = 0.002; MD: 3.0 min; 95% CI: 0.56–5.45).
Conclusions
The findings of this preliminary study seem to suggest that the tested prototype dCAIS system may be feasible to accurately place implants without conventional registration or optical tracking, potentially reducing surgical time. However, these findings should be interpreted with caution due to the study limitations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Sunshine Dental Clinic, Ningbo, China.
- Professor of the Department of Prosthodontics, College and Hospital of Stomatology, Guangxi Medical University, Nanning, China.
- Chief Engineer of the Department of Research and Development, Suzhou Digital-Health Care Co. Ltd., Suzhou, China.
- Full professor of Oral Surgery and Implantology, Director of the Department of Dentistry, Faculty of Medicine and Health Sciences, Universitat de Barcelona, Researcher of the IDIBELL Research Institute, Barcelona, Spain.
- Associate professor of Oral Surgery and Implantology, Faculty of Medicine and Health Sciences, Universitat de Barcelona, Researcher of the IDIBELL Research Institute, Barcelona, Spain.
ORCID-Profile: Eduard Valmaseda-Castellón, Rui Figueiredo
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
He, G., Liao, H., Sheng, H., Valmaseda-Castellón, E., & Figueiredo, R. (2026). Is Accurate Dental Implant Placement Feasible Using a Novel Dynamic Computer‐Assisted Surgery System Without Patient Optical Markers or Registration? A Preliminary Retrospective Cohort Study. Clinical and Experimental Dental Research. https://doi.org/10.1002/cre2.70256
Vancouver
He G, Liao H, Sheng H, Valmaseda-Castellón E, Figueiredo R. Is Accurate Dental Implant Placement Feasible Using a Novel Dynamic Computer‐Assisted Surgery System Without Patient Optical Markers or Registration? A Preliminary Retrospective Cohort Study. Clinical and Experimental Dental Research. 2026. doi:10.1002/cre2.70256
BibTeX
@article{He2026,
author = {He, Gang and Liao, Hongbing and Sheng, Hong and Valmaseda-Castellón, Eduard and Figueiredo, Rui},
title = {Is Accurate Dental Implant Placement Feasible Using a Novel Dynamic Computer‐Assisted Surgery System Without Patient Optical Markers or Registration? A Preliminary Retrospective Cohort Study},
journal = {Clinical and Experimental Dental Research},
year = {2026},
doi = {10.1002/cre2.70256}
}
EUVIMED