A novel contact force sensing pulsed field ablation catheter in a porcine model
Clinical Cardiology · 2024-02 · 47 (2)
DOI: 10.1002/clc.24220 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Cardiology, online erschienen am 29.01.2024. Verfasst von Juan Hua und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, The Second Affiliated Hospital of Nanchang University, Nanchang, China. Hauptthemen laut MEDLINE: Vena Cava, Superior, Catheter Ablation. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The Key Research and Development Program Project of Jiangxi province, China, The Science and Technology Innovation Action Plan of Shanghai, China.
Abstract (Originalsprache)
Background
Pulsed field ablation (PFA) has emerged as a novel non‐thermal modality with highly myocardium‐specific. However, the PFA catheter based on contact force (CF)‐sensing has not been reported. The study aimed to evaluate the efficacy and safety of a novel CF‐sensing PFA catheter.
Methods
First, different CF (5, 15, 25, and 35 g) of the novel PFA catheter were evaluated on lesion dimensions during ablation on right and left ventricle in two pigs. Next, this catheter was further evaluated on four typical sites of superior vena cava (SVC), cavotricuspid isthmus (CTI), right superior pulmonary vein (RSPV), and right inferior pulmonary vein (RIPV) for atrial ablation in another six pigs. Electrical isolation was evaluated immediately after ablation and 30‐day survival. Chronic lesions were assessed via histopathology after euthanasia. Acute and chronic safety outcomes were observed peri‐ and post‐procedurally.
Results
In ventricular ablation, increased CF from 5 to 15 g produced significantly greater lesion depth but nonsignificant increases from 15 to 35 g. In atrial ablation, the novel CF‐sensing PFA deliveries produced an acute attenuation of local electrograms and formation of a continuous line of block in all 6 pigs. The ablation line remained sustained blockage at the 30‐day survival period. The CF of SVC, CTI, RSPV, and RIPV was 9.4 ± 1.5, 14.5 ± 3.2, 17.2 ± 2.6, and 13.4 ± 2.8 g, respectively. Moreover, no evidence of damage to esophagus or phrenic nerve was observed.
Conclusion
The novel CF‐sensing PFA catheter potentiated efficient, safe, and durable ablation, without causing damage to the esophagus or phrenic nerve.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, The Second Affiliated Hospital of Nanchang University, Nanchang, China.
- Cardiac Electrophysiology R&D Center, APT Medical Inc., Shanghai, China.
Förderung
- The Key Research and Development Program Project of Jiangxi province, China
- The Science and Technology Innovation Action Plan of Shanghai, China
Zitieren
APA
Hua, J., Xiong, Q., Kong, Q., Xiong, L., Huang, Q., Hu, J., Li, J., Hu, J., Si, P., Zhou, T., & Chen, Q. (2024). A novel contact force sensing pulsed field ablation catheter in a porcine model. Clinical Cardiology, 47(2). https://doi.org/10.1002/clc.24220
Vancouver
Hua J, Xiong Q, Kong Q, Xiong L, Huang Q, Hu J, et al. A novel contact force sensing pulsed field ablation catheter in a porcine model. Clinical Cardiology. 2024;47(2). doi:10.1002/clc.24220
BibTeX
@article{Hua2024,
author = {Hua, Juan and Xiong, Qinmei and Kong, Qiling and Xiong, Liang and Huang, Qianghui and Hu, Jinzhu and Li, Juxiang and Hu, Jianxin and Si, Peng and Zhou, Tuo and Chen, Qi},
title = {A novel contact force sensing pulsed field ablation catheter in a porcine model},
journal = {Clinical Cardiology},
year = {2024},
volume = {47},
number = {2},
doi = {10.1002/clc.24220}
}
EUVIMED