EUVIMEDOpdag EUVIMED
Fallbericht

Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR

Ryotaro Iwata iD, Daisuke Fukamachi iD, Naoya Matsumoto iD, Yasuo Okumura iD

Catheterization and Cardiovascular Interventions · 2026-11-11

DOI: 10.1002/ccd.71020 ↗

At a glance

Fallbericht in Catheterization and Cardiovascular Interventions, published online on 2026-11-11. Written by Ryotaro Iwata and 3 co-authors. Institution: Division of Cardiology, Department of Medicine, Nihon University Hospital, Tokyo, Japan. 7 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Deployment‐balloon rupture during balloon‐expandable transcatheter aortic valve replacement can impede device retrieval. An octogenarian woman underwent transfemoral implantation of a 23‐mm SAPIEN 3 Ultra RESILIA valve. The balloon ruptured at approximately the nominal inflation volume adjacent to focal sinotubular‐junction calcification. Immediate aspiration produced complete collapse without residual contrast or abnormal balloon geometry, permitting resistance‐free retrieval into the 14‐F eSheath. Inspection demonstrated a longitudinal tear involving nearly the entire balloon length without circumferential separation or missing fragments. Tear orientation was identifiable only after retrieval; during the procedure, residual contrast, balloon geometry, response to aspiration, and withdrawal resistance guided management. This case suggests that rupture configuration may influence balloon collapse and retrievability, although the proposed relationship remains hypothesis‐generating.

Subjects

Author keywords

SAPIEN 3 Ultra RESILIATAVRballoon rupturedevice retrievalsinotubular junction

Authors and institutions

  • Division of Cardiology, Department of Medicine, Nihon University Hospital, Tokyo, Japan.
  • Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.

ORCID profiles: Ryotaro Iwata, Daisuke Fukamachi, Naoya Matsumoto, Yasuo Okumura

Cite

APA

Iwata, R., Fukamachi, D., Matsumoto, N., & Okumura, Y. (2026). Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.71020

Vancouver

Iwata R, Fukamachi D, Matsumoto N, Okumura Y. Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.71020

BibTeX
@article{Iwata2026,
  author = {Iwata, Ryotaro and Fukamachi, Daisuke and Matsumoto, Naoya and Okumura, Yasuo},
  title = {Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2026},
  doi = {10.1002/ccd.71020}
}
Start for free →