EUVIMEDEUVIMED entdecken
Fallbericht

Reverse Balloon Anchoring for Retrieval of an Entrapped Side‐Branch Balloon During the Jailed Balloon Technique

Takashi Sasaki iD, Kazuhide Ando iD, Hiroaki Kobayashi iD, Nogiku Niwamae iD

Catheterization and Cardiovascular Interventions · 09.11.2026

DOI: 10.1002/ccd.71013 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 09.11.2026. Verfasst von Takashi Sasaki und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Fukaya Red Cross Hospital, Fukaya, Japan. Die Arbeit nennt 10 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The jailed balloon technique (JBT) is an effective strategy for side branch (SB) protection during percutaneous coronary intervention (PCI) for coronary bifurcation lesions; however, jailed balloon entrapment is a rare but serious complication. We report the case of a man in his 70s who was successfully resuscitated after out‐of‐hospital cardiac arrest due to ventricular fibrillation and underwent emergency PCI for a right coronary artery bifurcation lesion. A drug‐eluting stent was deployed from the mid right coronary artery to the posterior descending branch using JBT. The jailed SB balloon became entrapped and could not be retrieved despite repeated inflation and deflation, traction, and rotational torque. Further traction resulted in deep engagement of the guiding catheter and an accordion phenomenon of the right coronary artery, raising concern about vascular injury. A balloon was therefore inflated within the stented main vessel, and forward pressure was applied to its shaft to provide counterforce against traction on the entrapped SB balloon. Using this technique, which we termed the “reverse balloon anchoring technique,” the entrapped balloon was successfully retrieved without apparent device damage. Final angiography and intravascular ultrasound demonstrated preserved SB flow and adequate stent apposition. The reverse balloon anchoring technique may represent a useful bailout option for jailed balloon entrapment when conventional retrieval maneuvers fail.

Schlagwörter

Stichwörter der Autor:innen

balloon anchoringballoon entrapmentbifurcation lesionjailed balloon techniquepercutaneous coronary intervention

Autor:innen und Institutionen

  • Department of Cardiology, Fukaya Red Cross Hospital, Fukaya, Japan.
  • Department of Cardiovascular Medicine, Red Cross Maebashi Hospital, Maebashi, Japan.

ORCID-Profile: Takashi Sasaki, Kazuhide Ando, Hiroaki Kobayashi, Nogiku Niwamae

Zitieren

APA

Sasaki, T., Ando, K., Kobayashi, H., & Niwamae, N. (2026). Reverse Balloon Anchoring for Retrieval of an Entrapped Side‐Branch Balloon During the Jailed Balloon Technique. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.71013

Vancouver

Sasaki T, Ando K, Kobayashi H, Niwamae N. Reverse Balloon Anchoring for Retrieval of an Entrapped Side‐Branch Balloon During the Jailed Balloon Technique. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.71013

BibTeX
@article{Sasaki2026,
  author = {Sasaki, Takashi and Ando, Kazuhide and Kobayashi, Hiroaki and Niwamae, Nogiku},
  title = {Reverse Balloon Anchoring for Retrieval of an Entrapped Side‐Branch Balloon During the Jailed Balloon Technique},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2026},
  doi = {10.1002/ccd.71013}
}
Kostenlos starten →