EUVIMEDObjevte EUVIMED
VergleichsstudieOpen access

Outcomes of Different Device Sizing Strategies Using the Newer‐Generation WATCHMAN Devices for Closure of Borderline‐Sized Left Atrial Appendage

Motoki Fukutomi, Naoki Saito, Shinya Wada, Mizuho Hoshina, Shunsuke Sasaki, Sunao Nakamura iD

Catheterization and Cardiovascular Interventions · 2026

DOI: 10.1002/ccd.70706 ↗

At a glance

Vergleichsstudie in Catheterization and Cardiovascular Interventions, 2026. Written by Motoki Fukutomi and 5 co-authors. Institution: Department of Cardiology, Kawasaki Saiwai Hospital, Kawasaki, Japan. Major MEDLINE topics: Left Atrial Appendage Closure, Atrial Fibrillation, Atrial Appendage. 19 references. The full text is freely available in PubMed Central.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

In left atrial appendage closure (LAAC) using newer‐generation WATCHMAN FLX/FLX Pro devices, managing borderline‐sized LAA within overlapping zones is common, but optimal device sizing strategies remain unknown.

Aims

To compare clinical outcomes and device‐related complications between smaller and larger device sizing strategies for borderline‐sized LAA.

Methods

This single‐center retrospective study evaluated 91 patients undergoing newer‐generation WATCHMAN implantation for borderline‐sized LAA between September 2021 and June 2024. Follow‐up imaging (transesophageal echocardiography and/or cardiac computed tomography angiography) was performed 45‐120 days post‐LAAC, and one‐year clinical outcomes were assessed.

Results

Patients were divided into smaller ( n = 36) and larger ( n = 55) device groups. The composite primary endpoint (cardiovascular mortality and ischemic stroke) was similar between groups (8.3% vs. 3.6%; hazard ratio [HR], 0.372; 95% CI: 0.074–2.650, p = 0.334). However, the secondary efficacy endpoint‐peridevice leak (PDL) 〉3 mm and/or device‐related thrombus (DRT)‐was significantly higher in the smaller device group at initial follow‐up, even after covariate adjustment (13.9% vs. 1.8%; HR, 0.091; 95% CI: 0.009–0.887, p = 0.039). Similar findings were observed at one year.

Conclusion

For borderline‐sized LAA treated with newer‐generation WATCHMAN devices, smaller and larger sizing strategies yielded similar one‐year clinical outcomes. However, the larger device strategy showed a lower incidence of PDL >3 mm and/or DRT.

Subjects

MeSH terms (MEDLINE)

Left Atrial Appendage ClosureAtrial FibrillationAtrial AppendageHumansRetrospective StudiesProsthesis DesignTreatment OutcomeTime FactorsMaleFemaleAgedRisk FactorsRisk AssessmentAged, 80 and overEchocardiography, TransesophagealMiddle Aged

Author keywords

borderline‐sized left atrial appendagedevice‐related thrombuslarger deviceleft atrial appendage closureperi‐device leak

Authors and institutions

  • Department of Cardiology, Kawasaki Saiwai Hospital, Kawasaki, Japan.

ORCID profiles: Sunao Nakamura

Cite

APA

Fukutomi, M., Saito, N., Wada, S., Hoshina, M., Sasaki, S., & Nakamura, S. (2026). Outcomes of Different Device Sizing Strategies Using the Newer‐Generation WATCHMAN Devices for Closure of Borderline‐Sized Left Atrial Appendage. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70706

Vancouver

Fukutomi M, Saito N, Wada S, Hoshina M, Sasaki S, Nakamura S. Outcomes of Different Device Sizing Strategies Using the Newer‐Generation WATCHMAN Devices for Closure of Borderline‐Sized Left Atrial Appendage. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70706

BibTeX
@article{Fukutomi2026,
  author = {Fukutomi, Motoki and Saito, Naoki and Wada, Shinya and Hoshina, Mizuho and Sasaki, Shunsuke and Nakamura, Sunao},
  title = {Outcomes of Different Device Sizing Strategies Using the Newer‐Generation WATCHMAN Devices for Closure of Borderline‐Sized Left Atrial Appendage},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2026},
  doi = {10.1002/ccd.70706}
}
Start for free →