Vergleichsstudie in der Zeitschrift Catheterization and Cardiovascular Interventions, 2026. Verfasst von Motoki Fukutomi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Kawasaki Saiwai Hospital, Kawasaki, Japan. Hauptthemen laut MEDLINE: Left Atrial Appendage Closure, Atrial Fibrillation, Atrial Appendage. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
In left atrial appendage closure (LAAC) using newer‐generation WATCHMAN FLX/FLX Pro devices, managing borderline‐sized LAA within overlapping zones is common, but optimal device sizing strategies remain unknown.
Aims
To compare clinical outcomes and device‐related complications between smaller and larger device sizing strategies for borderline‐sized LAA.
Methods
This single‐center retrospective study evaluated 91 patients undergoing newer‐generation WATCHMAN implantation for borderline‐sized LAA between September 2021 and June 2024. Follow‐up imaging (transesophageal echocardiography and/or cardiac computed tomography angiography) was performed 45‐120 days post‐LAAC, and one‐year clinical outcomes were assessed.
Results
Patients were divided into smaller ( n = 36) and larger ( n = 55) device groups. The composite primary endpoint (cardiovascular mortality and ischemic stroke) was similar between groups (8.3% vs. 3.6%; hazard ratio [HR], 0.372; 95% CI: 0.074–2.650, p = 0.334). However, the secondary efficacy endpoint‐peridevice leak (PDL) 〉3 mm and/or device‐related thrombus (DRT)‐was significantly higher in the smaller device group at initial follow‐up, even after covariate adjustment (13.9% vs. 1.8%; HR, 0.091; 95% CI: 0.009–0.887, p = 0.039). Similar findings were observed at one year.
Conclusion
For borderline‐sized LAA treated with newer‐generation WATCHMAN devices, smaller and larger sizing strategies yielded similar one‐year clinical outcomes. However, the larger device strategy showed a lower incidence of PDL >3 mm and/or DRT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Kawasaki Saiwai Hospital, Kawasaki, Japan.
ORCID-Profile: Sunao Nakamura
Zitieren
APA
Fukutomi, M., Saito, N., Wada, S., Hoshina, M., Sasaki, S., & Nakamura, S. (2026). Outcomes of Different Device Sizing Strategies Using the Newer‐Generation WATCHMAN Devices for Closure of Borderline‐Sized Left Atrial Appendage. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70706
Vancouver
Fukutomi M, Saito N, Wada S, Hoshina M, Sasaki S, Nakamura S. Outcomes of Different Device Sizing Strategies Using the Newer‐Generation WATCHMAN Devices for Closure of Borderline‐Sized Left Atrial Appendage. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70706
BibTeX
@article{Fukutomi2026,
author = {Fukutomi, Motoki and Saito, Naoki and Wada, Shinya and Hoshina, Mizuho and Sasaki, Shunsuke and Nakamura, Sunao},
title = {Outcomes of Different Device Sizing Strategies Using the Newer‐Generation WATCHMAN Devices for Closure of Borderline‐Sized Left Atrial Appendage},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
doi = {10.1002/ccd.70706}
}
EUVIMED