EUVIMEDOpdag EUVIMED
Fallbericht

Harmony Transcatheter Pulmonary Valve Housing Compression: Bench Testing and Case Report

Mary Breige O'Donnell-Smith, Daniel S Levi, Gentian Lluri, Jamil Aboulhosn

Catheterization and Cardiovascular Interventions · 2026

DOI: 10.1002/ccd.70382 ↗

At a glance

Fallbericht in Catheterization and Cardiovascular Interventions, 2026. Written by Mary Breige O'Donnell-Smith and 3 co-authors. Institution: Division of Pediatric Cardiology, University of Washington School of Medicine, Seattle Children's Hospital, Seattle, Washington, USA. Major MEDLINE topics: Heart Valve Prosthesis, Cardiac Catheterization, Heart Valve Prosthesis Implantation, Bioprosthesis. 4 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Use of the Harmony valve within a failing surgical bioprosthesis has not been described to date, especially with respect to compression of the valve and valve housing when Harmony is used in smaller, non‐compliant landing zones. Described is a case of a 45‐year‐old male with a history of congenital pulmonary valve stenosis who underwent surgical pulmonary valve placement with a 25‐mm Biocor Epic bioprosthesis and developed pulmonary regurgitation. The patient underwent cardiac catheterization with high‐pressure balloon valvuloplasty with intentional fracturing and resultant compression of the left anterior descending (LAD) coronary artery. Hence, a balloon‐expandable valve platform was not implanted due to concern for coronary artery compression. A bench assessment with flow analysis was performed to determine the suitability of the Harmony valve in this setting using ex vivo hydrodynamic testing with circular and elliptical compressions of the Harmony valve. Patient‐specific testing was performed with additional fixture sizes created and tested to better understand valve performance under differing levels of compression. The tested size range for TPV22 and TPV25 started at 19 mm and increased every 1 mm in diameter until the unconstrained valve housing was reached. After bench testing, the patient underwent implantation of Harmony TPV25 with excellent immediate and intermediate‐term outcomes.

Subjects

MeSH terms (MEDLINE)

Heart Valve ProsthesisCardiac CatheterizationHeart Valve Prosthesis ImplantationBioprosthesisPulmonary Valve InsufficiencyProsthesis FailurePulmonary ValvePulmonary Valve StenosisHumansMaleMiddle AgedProsthesis DesignHemodynamicsTreatment OutcomeBalloon ValvuloplastyMaterials Testing

Author keywords

RVOTbioprosthetic valvecongenital heart diseasepulmonary regurgitation

Authors and institutions

  • Division of Pediatric Cardiology, University of Washington School of Medicine, Seattle Children's Hospital, Seattle, Washington, USA.
  • Division of Pediatric Cardiology, David Geffen School of Medicine at UCLA, Los Angeles, California, USA.
  • Ahmanson/UCLA Adult Congenital Heart Center, David Geffen School of Medicine at UCLA, Los Angeles, California, USA.

Cite

APA

O'Donnell-Smith, M. B., Levi, D. S., Lluri, G., & Aboulhosn, J. (2026). Harmony Transcatheter Pulmonary Valve Housing Compression: Bench Testing and Case Report. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70382

Vancouver

O'Donnell-Smith MB, Levi DS, Lluri G, Aboulhosn J. Harmony Transcatheter Pulmonary Valve Housing Compression: Bench Testing and Case Report. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70382

BibTeX
@article{ODonnellSmith2026,
  author = {O'Donnell-Smith, Mary Breige and Levi, Daniel S and Lluri, Gentian and Aboulhosn, Jamil},
  title = {Harmony Transcatheter Pulmonary Valve Housing Compression: Bench Testing and Case Report},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2026},
  doi = {10.1002/ccd.70382}
}
Start for free →