Commissural Alignment in SAPIEN 3 Valves: Impact on Gradient and Mortality at Follow‐Up
Catheterization and Cardiovascular Interventions · 2025
DOI: 10.1002/ccd.31418 ↗At a glance
Journal article in Catheterization and Cardiovascular Interventions, 2025. Written by Leire Unzue and 9 co-authors. Institution: Departamento de Cardiología Intervencionista, HM CIEC MADRID (Centro Integral de Enfermedades Cardiovasculares), HM Hospitales, Madrid, Spain. Major MEDLINE topics: Heart Valve Prosthesis, Aortic Valve, Aortic Valve Stenosis, Hemodynamics. Cited 1 times so far (Crossref); 31 references.
Abstract (original language)
Background
Commissural misalignment (CMA) has demonstrated to difficult future coronary access in self‐expandable prosthesis and has been related with subclinical leaflet thrombosis and a potential increase in valve gradients and aortic regurgitation at follow‐up.
Aims
The aim of our study was to assess the commissural alignment of the balloon expandable SAPIEN 3 valve (Edwards Lifesciences) and to evaluate its impact on transvalvular gradient and mortality at 5 years of follow‐up.
Methods
Consecutive patients treated with SAPIEN 3 prosthesis guided with intraprocedural transesophageal echocardiogram (TEE) with a minimal follow‐up of 5 years were included. Commissural alignment between the prosthesis and the native valve was measured by TEE by two independent cardiologists using a standardized 3‐cusp view pre and post‐valve deployment. CMA was defined as a neocommissure deviation > 30° compared with native commissures. Echocardiographic gradients and clinical events were recorded during follow‐up and compared between groups.
Results
One hundred and five patients (54% male, mean age 80.4 years) were included; 63 (40%) were classified as “aligned” (CMA < 30°) and 42 (60%) as “misaligned” (CMA > 30°). Patients with CMA > 30° presented smaller aortic annulus (23.8 vs. 20.7, p < 0.01) and received smaller valves. There was a slight increase of mean transvalvular gradient during follow‐up (9.2−10.3 mmHg, p < 0.05), without significant differences between groups. Mortality rate was 16.2% at 1 year and 49.5% at 5 years; and not related to commissural alignment.
Conclusions
Significant CMA is present in 40% of patients treated with SAPIEN 3 valves, however, we didn't find a significant association between CMA and gradient increase or mortality during long‐term follow‐up.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Departamento de Cardiología Intervencionista, HM CIEC MADRID (Centro Integral de Enfermedades Cardiovasculares), HM Hospitales, Madrid, Spain.
- Instituto de Investigación Sanitaria, HM Hospitales, Madrid, Spain.
- Departamento de imagen cardiaca, HM CIEC MADRID (Centro Integral de Enfermedades Cardiovasculares), HM Hospitales, Madrid, Spain.
- Facultad HM de Ciencias de la Salud de la Universidad Camilo José Cela, Madrid, Spain.
- Centro de Investigación en Red de Enfermedades Respiratorias (CIBERES), Instituto de Salud Carlos III (ISCIII), Madrid, Spain.
- Departamento de Cardiología, HM CIEC MADRID (Centro Integral de Enfermedades Cardiovasculares), HM Hospitales, Madrid, Spain.
ORCID profiles: Leire Unzue
Cite
APA
Unzue, L., Antón, B. D., Laderas, A. P., Polo, L. M., Vázquez, J. M. C., Rodrigo, F. J. R., Jimenez, F. J. P., Fernández-Friera, L., Mestre, R. T., & Fernández, E. G. (2025). Commissural Alignment in SAPIEN 3 Valves: Impact on Gradient and Mortality at Follow‐Up. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31418
Vancouver
Unzue L, Antón BD, Laderas AP, Polo LM, Vázquez JMC, Rodrigo FJR, et al. Commissural Alignment in SAPIEN 3 Valves: Impact on Gradient and Mortality at Follow‐Up. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31418
BibTeX
@article{Unzue2025,
author = {Unzue, Leire and Antón, Belén Díaz and Laderas, Adrián Peláez and Polo, Lorena Martín and Vázquez, Jose María Castellano and Rodrigo, Francisco José Rodríguez and Jimenez, Francisco Javier Parra and Fernández-Friera, Leticia and Mestre, Rodrigo Teijeiro and Fernández, Eulogio García},
title = {Commissural Alignment in SAPIEN 3 Valves: Impact on Gradient and Mortality at Follow‐Up},
journal = {Catheterization and Cardiovascular Interventions},
year = {2025},
doi = {10.1002/ccd.31418}
}
EUVIMED