EUVIMEDDescoperă EUVIMED
Systematische ÜbersichtMetaanalyse

Systematic Review and Meta‐Analysis of the Super High‐Pressure Balloon (SIS‐OPN) for Percutaneous Coronary Intervention

Sant Kumar iD, Sandeep Jalli, Yader Sandoval iD, Khaldoon Alaswad, Nachiket J Patel iD, Tim Henry iD, Darshan Doshi, Ashish Pershad, Arnold Seto, Gautam Kumar iD, M Nicholas Burke, Emmanouil S Brilakis iD

Catheterization and Cardiovascular Interventions · 2025

DOI: 10.1002/ccd.31403 ↗

At a glance

Systematische Übersicht in Catheterization and Cardiovascular Interventions, 2025. Written by Sant Kumar and 11 co-authors. Institution: Department of Cardiology, Creighton University School of Medicine, Phoenix, Arizona, USA. Major MEDLINE topics: Coronary Artery Disease, Angioplasty, Balloon, Coronary, Cardiac Catheters, Percutaneous Coronary Intervention. Cited 2 times so far (Crossref); 36 references. Funded by: The authors received no specific funding for this work..

RezumatSubjectsDate bibliograficeCiteRelated studies

Abstract (original language)

Introduction

The super high‐pressure NC balloon (OPN NC; SIS Medical AG, Winterthur, Switzerland) is increasingly used in percutaneous coronary intervention (PCI). We performed a systematic review and meta‐analysis of its efficacy and safety.

Methods

A systematic review was conducted using PubMed and the Cochrane Library to identify studies using the OPN NC balloon in PCI. Procedural success, major adverse cardiac events (MACE), coronary perforation, and other complications were pooled using weighted means, confidence intervals (CI), and I ² statistics for heterogeneity assessment.

Results

Twenty‐nine studies (5 prospective including 2 randomized‐controlled trials, 12 retrospective, and 12 case reports) were included in the systematic review, of which 11 (3 prospective, 8 retrospective) were included in the meta‐analysis. Of the 1015 meta‐analysis patients, 50.7% presented with stable angina and 21.6% with an acute coronary syndrome. Mean lesion length was 22.8 ± 13.7 mm, the prevalence of moderate or severe calcification was 93.7%, and 40.4% of lesions were in‐stent. Procedural success was 95.1% (95% CI 89.3%−98.8% with significant heterogeneity: I ² = 84.9%). The incidence of periprocedural MACE was 1.9% (95% CI 1.1%−2.8%, I ² 86.9%), and the incidence of coronary perforation was 0.6% (95% CI 0.3%−1.5%, I ² 87.2%).

Conclusions

Use of the OPN NC balloon is associated with high success and acceptable complication rates. The heterogeneity of outcomes underscores the need for additional studies and standardized definitions.

Subjects

MeSH terms (MEDLINE)

Coronary Artery DiseaseAngioplasty, Balloon, CoronaryCardiac CathetersPercutaneous Coronary InterventionHumansTreatment OutcomeRisk FactorsEquipment DesignAgedMaleFemalePressureMiddle Aged

Author keywords

calcified lesionspercutaneous coronary interventionstent expansionsuper high‐pressure balloon

Authors and institutions

  • Department of Cardiology, Creighton University School of Medicine, Phoenix, Arizona, USA.
  • Minneapolis Heart Institute and Minneapolis Heart Institute Foundation, Abbott Northwestern Hospital, Minneapolis, Minnesota, USA.
  • Division of Cardiovascular Disease, Henry Ford Hospital, Detroit, Michigan, USA.
  • Department of Cardiology, Chandler Regional Medical Center, Chandler, Arizona, USA.
  • Department of Cardiology, The Carl and Edyth Lindner Center for Research and Education, Cincinnati, Ohio, USA.
  • Division of Cardiology, Massachusetts General Hospital, Boston, Massachusetts, USA.

ORCID profiles: Sant Kumar, Yader Sandoval, Nachiket J Patel, Tim Henry, Gautam Kumar, Emmanouil S Brilakis

Funding

  • The authors received no specific funding for this work.

Cite

APA

Kumar, S., Jalli, S., Sandoval, Y., Alaswad, K., Patel, N. J., Henry, T., Doshi, D., Pershad, A., Seto, A., Kumar, G., Burke, M. N., & Brilakis, E. S. (2025). Systematic Review and Meta‐Analysis of the Super High‐Pressure Balloon (SIS‐OPN) for Percutaneous Coronary Intervention. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31403

Vancouver

Kumar S, Jalli S, Sandoval Y, Alaswad K, Patel NJ, Henry T, et al. Systematic Review and Meta‐Analysis of the Super High‐Pressure Balloon (SIS‐OPN) for Percutaneous Coronary Intervention. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31403

BibTeX
@article{Kumar2025,
  author = {Kumar, Sant and Jalli, Sandeep and Sandoval, Yader and Alaswad, Khaldoon and Patel, Nachiket J and Henry, Tim and Doshi, Darshan and Pershad, Ashish and Seto, Arnold and Kumar, Gautam and Burke, M Nicholas and Brilakis, Emmanouil S},
  title = {Systematic Review and Meta‐Analysis of the Super High‐Pressure Balloon (SIS‐OPN) for Percutaneous Coronary Intervention},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2025},
  doi = {10.1002/ccd.31403}
}
Start for free →