Head‐to‐Head Comparison of Learning Curves Between QFR and FFRangio Software Users
Catheterization and Cardiovascular Interventions · 2025
DOI: 10.1002/ccd.31384 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Adil Salihu und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland. Hauptthemen laut MEDLINE: Learning Curve, Coronary Angiography, Coronary Stenosis, Fractional Flow Reserve, Myocardial. Die Arbeit nennt 13 Literaturangaben. Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Background
Quantitative flow ratio (QFR) and FFRangio are angiography‐based technologies used to perform functional assessment of coronary lesions from angiographic images, validated across multiple clinical studies. There is limited information on the learning curves associated with each technology.
Aims
This study aims to compare the learning curves of QFR and FFRangio in evaluating coronary stenoses, focusing on changes in analysis speed and accuracy compared to invasive measurements.
Methods
A team of five blinded investigators, including two nurses, one medical student, and one physician in training, underwent identical standardized training on both technologies. The time taken for each analysis and the computed FFR values were documented and compared against the invasive gold standard.
Results
A total of 270 lesions (54 coronary lesions in 44 patients) were retrospectively analyzed. The median invasive FFR value was 0.88 [IQR 0.5, 0.9]. The median time for analysis with QFR and FFRangio was 245 [IQR 62, 319] and 252 [IQR 82, 315] s, respectively ( p = 0.171). Both QFR and FFRangio demonstrated a significant reduction in the time required for analysis as experience increased ( p < 0.01). Regarding accuracy, the median difference with invasive FFR for QFR and FFRangio was 0.06 [IQR: 0, 0.12] and 0.06 [IQR: 0, 0.12], respectively ( p = 0.620). Both technologies reached a performance plateau early on, exhibiting comparable results throughout the study.
Conclusion
Initial training in QFR and FFRangio enables quick attainment of maximal performance, but further practice primarily enhances analysis speed while maintaining accuracy, for both software.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.
- MicroBioRobotic Systems Laboratory, Institute of Mechanical Engineering, EPFL, Lausanne, Switzerland.
ORCID-Profile: Adil Salihu, David Meier, Stephane Fournier
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Salihu, A., Zulauff, J., Gadiri, M. A., Metzinger, A., Muller, J., Skalidis, I., Meier, D., Noirclerc, N., Mauler-Wittwer, S., Zimmerli, A., Muller, O., & Fournier, S. (2025). Head‐to‐Head Comparison of Learning Curves Between QFR and FFRangio Software Users. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31384
Vancouver
Salihu A, Zulauff J, Gadiri MA, Metzinger A, Muller J, Skalidis I, et al. Head‐to‐Head Comparison of Learning Curves Between QFR and FFRangio Software Users. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31384