Finite element modeling with patient‐specific geometry to assess clinical risks of percutaneous pulmonary valve implantation

Carly L Donahue iD, Claire L Westman, Brittany L Faanes, Athar M Qureshi iD, Victor H Barocas, Varun Aggarwal

Catheterization and Cardiovascular Interventions · 2024

DOI: 10.1002/ccd.31016 ↗

At a glance

Journal article in Catheterization and Cardiovascular Interventions, 2024. Written by Carly L Donahue and 5 co-authors. Institution: Department of Biomedical Engineering, University of Minnesota, Twin Cities, Minnesota, USA. Major MEDLINE topics: Pulmonary Valve, Heart Valve Prosthesis Implantation, Finite Element Analysis, Patient-Specific Modeling. Cited 2 times so far (Crossref); 36 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: NIH grant U01-HL139471, T32-HL139431, Andrew David Sit Foundation Innovators Fund award.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Percutaneous pulmonary valve implantation (PPVI) is a non‐surgical treatment for right ventricular outflow tract (RVOT) dysfunction. During PPVI, a stented valve, delivered via catheter, replaces the dysfunctional pulmonary valve. Stent oversizing allows valve anchoring within the RVOT, but overexpansion can intrude on the surrounding structures. Potentially dangerous outcomes include aortic valve insufficiency (AVI) from aortic root (AR) distortion and myocardial ischemia from coronary artery (CA) compression. Currently, risks are evaluated via balloon angioplasty/sizing before stent deployment. Patient‐specific finite element (FE) analysis frameworks can improve pre‐procedural risk assessment, but current methods require hundreds of hours of high‐performance computation.

Methods

We created a simplified method to simulate the procedure using patient‐specific FE models for accurate, efficient pre‐procedural PPVI (using balloon expandable valves) risk assessment. The methodology was tested by retrospectively evaluating the clinical outcome of 12 PPVI candidates.

Results

Of 12 patients (median age 14.5 years) with dysfunctional RVOT, 7 had native RVOT and 5 had RV‐PA conduits. Seven patients had undergone successful RVOT stent/valve placement, three had significant AVI on balloon testing, one had left CA compression, and one had both AVI and left CA compression. A model‐calculated change of more than 20% in lumen diameter of the AR or coronary arteries correctly predicted aortic valve sufficiency and/or CA compression in all the patients.

Conclusion

Agreement between FE results and clinical outcomes is excellent. Additionally, these models run in 2–6 min on a desktop computer, demonstrating potential use of FE analysis for pre‐procedural risk assessment of PPVI in a clinically relevant timeframe.

Subjects

MeSH terms (MEDLINE)

Pulmonary ValveHeart Valve Prosthesis ImplantationFinite Element AnalysisPatient-Specific ModelingHeart Valve ProsthesisModels, CardiovascularProsthesis DesignCardiac CatheterizationHumansRisk AssessmentAdolescentTreatment OutcomeRisk FactorsMaleChildRetrospective StudiesFemaleYoung AdultPredictive Value of TestsHemodynamicsStentsPulmonary Valve InsufficiencyVentricular Outflow ObstructionClinical Decision-MakingAdult

Author keywords

balloon testingpulmonary valvetranscathetervalve replacement

Authors and institutions

  • Department of Biomedical Engineering, University of Minnesota, Twin Cities, Minnesota, USA.
  • Division of Pediatric Cardiology, Department of Pediatrics, University of Minnesota Masonic Children's Hospital, Minneapolis, Minnesota, USA.
  • Department of Pediatrics, The Lillei Frank Abercombie Section of Pediatric Cardiology, Texas Children's Hospital and Baylor College of Medicine, Houston, Texas, USA.

ORCID profiles: Carly L Donahue, Athar M Qureshi

Funding

  • NIH grant U01-HL139471
  • T32-HL139431
  • Andrew David Sit Foundation Innovators Fund award
  • NHLBI NIH HHS

Cite

APA

Donahue, C. L., Westman, C. L., Faanes, B. L., Qureshi, A. M., Barocas, V. H., & Aggarwal, V. (2024). Finite element modeling with patient‐specific geometry to assess clinical risks of percutaneous pulmonary valve implantation. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31016

Vancouver

Donahue CL, Westman CL, Faanes BL, Qureshi AM, Barocas VH, Aggarwal V. Finite element modeling with patient‐specific geometry to assess clinical risks of percutaneous pulmonary valve implantation. Catheterization and Cardiovascular Interventions. 2024. doi:10.1002/ccd.31016

BibTeX
@article{Donahue2024,
  author = {Donahue, Carly L and Westman, Claire L and Faanes, Brittany L and Qureshi, Athar M and Barocas, Victor H and Aggarwal, Varun},
  title = {Finite element modeling with patient‐specific geometry to assess clinical risks of percutaneous pulmonary valve implantation},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2024},
  doi = {10.1002/ccd.31016}
}
Start for free →