EUVIMEDEUVIMED entdecken
Open Access

Image‐Derived Modeling to Assess Coronary Proximity in Patients Undergoing Transcatheter Pulmonary Valve Replacement With Self‐Expanding Valves

Yuval Barak-Corren, Vladislav Obsekov, Mudit Gupta iD, Christian Herz, Silvani Amin, Andras Lasso, Michael L O'Byrne iD, Matthew J Gillespie, Matthew A Jolley iD

Catheterization and Cardiovascular Interventions · 2025

DOI: 10.1002/ccd.31469 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Yuval Barak-Corren und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatrics, Division of Pediatric Cardiology, The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Pulmonary Valve, Cardiac Catheterization, Heart Valve Prosthesis Implantation, Heart Valve Prosthesis. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NHLBI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The potential for coronary artery compression (CC) during transcatheter pulmonary valve replacement (TPVR) using self‐expanding valves (SEV) is not fully understood, yet anecdotal reports suggest that this risk exists. Aims and Methods We performed a retrospective cohort study of patients evaluated for SEV‐TPVR to evaluate the relationship between the right ventricular outflow tract (RVOT) and coronary arteries (CA). CT‐derived segmentations of the RVOT and CA were created using machine learning. A 2D map of the distance between the RVOT surface and CA, in systole and diastole, was created. In the subset of patients with post‐procedural CTA, the distance before and after TPVR was measured.

Results

Forty‐two individuals underwent screening for SEV‐TPVR, of which 83% ( n = 35) had SEV implanted (Harmony = 24; Alterra = 11). Median age was 22.9 years (range 12−60) and 76% had tetralogy of Fallot (TOF). There was no significant change in the distance between the RVOT and LCA between diastole and systole ( p = 0.31), yet the RVOT area nearest to the LCA displaced proximally by 11 mm (IQR: 5.6−19.9) in systole. In 8 patients with pre‐ and post‐TPVR CTA, no statistically significant differences were observed in the RVOT‐to‐LCA relation after intervention. The distance to the LCA was smaller in pulmonary stenosis/atresia patients than those with TOF (median distance 1.2 and 2.1 mm, respectively; p = 0.185).

Conclusion

The RVOT area in closest proximity to LCA is dynamic and should be considered when planning TPVR. Special attention should be given to patients with a diagnosis of pulmonary stenosis/atresia.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pulmonary ValveCardiac CatheterizationHeart Valve Prosthesis ImplantationHeart Valve ProsthesisComputed Tomography AngiographyCoronary VesselsCoronary AngiographyPatient-Specific ModelingRadiographic Image Interpretation, Computer-AssistedMultidetector Computed TomographyHumansRetrospective StudiesChildMaleFemaleAdolescentAdultYoung AdultTreatment OutcomeMiddle AgedPredictive Value of TestsProsthesis DesignPulmonary Valve InsufficiencyMachine LearningRisk FactorsHemodynamics

Stichwörter der Autor:innen

3D modelingcoronary artery compression (CC)machine learningself‐expanding valves (SEV)transcatheter pulmonary valve replacement (TPVR)

Autor:innen und Institutionen

  • Department of Pediatrics, Division of Pediatric Cardiology, The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA.
  • Division of Pediatric Cardiology, The Mount Sinai Hospital, New York, New York, USA.
  • Hospital of the University of Pennsylvania, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Department of Anesthesiology and Critical Care Medicine, The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA.
  • Laboratory for Percutaneous Surgery, Queen's University, Kingston, Ontario, CA.

ORCID-Profile: Mudit Gupta, Michael L O'Byrne, Matthew A Jolley

Förderung

  • NHLBI NIH HHS

Zitieren

APA

Barak-Corren, Y., Obsekov, V., Gupta, M., Herz, C., Amin, S., Lasso, A., O'Byrne, M. L., Gillespie, M. J., & Jolley, M. A. (2025). Image‐Derived Modeling to Assess Coronary Proximity in Patients Undergoing Transcatheter Pulmonary Valve Replacement With Self‐Expanding Valves. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31469

Vancouver

Barak-Corren Y, Obsekov V, Gupta M, Herz C, Amin S, Lasso A, et al. Image‐Derived Modeling to Assess Coronary Proximity in Patients Undergoing Transcatheter Pulmonary Valve Replacement With Self‐Expanding Valves. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31469

BibTeX
@article{BarakCorren2025,
  author = {Barak-Corren, Yuval and Obsekov, Vladislav and Gupta, Mudit and Herz, Christian and Amin, Silvani and Lasso, Andras and O'Byrne, Michael L and Gillespie, Matthew J and Jolley, Matthew A},
  title = {Image‐Derived Modeling to Assess Coronary Proximity in Patients Undergoing Transcatheter Pulmonary Valve Replacement With Self‐Expanding Valves},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2025},
  doi = {10.1002/ccd.31469}
}
Kostenlos starten →