EUVIMEDScopri EUVIMED
Open access · CC BY-NC 4.0

Single center experience with the balloon‐expandable Myval transcatheter aortic valve system with the first 100 patients: 30‐day and 1‐year follow‐up

Balázs Magyari iD, Bálint Kittka, Ilona Goják, Gábor Kasza, Kristóf Schönfeld, László Botond Szapáry, Mihály Simon, Rudolf Kiss, Andrea Bertalan, Edit Várady, András Gyimesi, István Szokodi and 1 more

Catheterization and Cardiovascular Interventions · 2023

DOI: 10.1002/ccd.30868 ↗

At a glance

Journal article in Catheterization and Cardiovascular Interventions, 2023. Written by Balázs Magyari and 12 co-authors. Institution: Heart Institute, Medical School, University of Pécs, Pécs, Hungary. Major MEDLINE topics: Aortic Valve Stenosis, Transcatheter Aortic Valve Replacement, Heart Valve Prosthesis. Cited 9 times so far (Crossref); 34 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: National Research, Development, and Innovation Office of Hungary.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Aims

To report our single‐center data, regarding the first 100 patients who underwent TAVR procedure with the new balloon‐expandable MYVAL system. We report 30‐day and 1‐year outcomes in low to high‐risk TAVR patient population.

Methods

From November 2019 to July 2021, 100 consecutive patients underwent TAVR procedure. Patient outcome was classified according to the VARC‐2 definitions. The device performance was assessed using transthoracic echocardiography. Data collection was allowed by the Local Ethical Committee.

Results

The mean age was 74.7 years, 63 (63%) were male. The mean Euroscore II and STS score were 4.8 ± 4.9 and 5.6 ± 3.9, respectively. Transfemoral access was the most frequent (surgical vs. percutaneous 2% vs. 97%) and in one patient surgical subclavian access was used. VARC‐2 outcomes were as follows: device success 99%, STROKE 1%, major and minor vascular complication was 1% and 11%, respectively, the rate of new permanent pacemaker implantation was 30.7%. At discharge, the incidence of grade I, grade II aortic regurgitation was 39% and 1%, respectively, without relevant PVL. In‐hospital mortality was only 1%. These results included a high proportion (17%) of patients with bicuspid aortic valves. At 1 year, the all‐cause mortality rate was 7% (only two due to cardiac event) and only a single patient had valve‐related dysfunction requiring surgical aortic replacement.

Conclusions

TAVR procedure with MYVAL transcatheter heart valve system shows excellent 30‐day and 1‐year outcomes regarding patient survival, technical success, and valve‐related adverse events. The limitations of our study comprise a single‐center study with retrospective data collection.

Subjects

MeSH terms (MEDLINE)

Aortic Valve StenosisTranscatheter Aortic Valve ReplacementHeart Valve ProsthesisHumansMaleAgedFemaleAortic ValveFollow-Up StudiesRetrospective StudiesTreatment OutcomeProsthesis Design

Author keywords

TAVRannular ruptureballoon expendable transcatheter heart valvebicuspid aortic valveparavalvular leakpermanent pacemaker implantation

Authors and institutions

  • Heart Institute, Medical School, University of Pécs, Pécs, Hungary.
  • Szentágothai Research Centre, University of Pécs, Pécs, Hungary.
  • Department of Vascular Surgery, Medical School, University of Pécs, Pécs, Hungary.
  • Department of Medical Imaging, Medical School, University of Pécs, Pécs, Hungary.
  • EconNet Research Group, Faculty of Business and Economics, University of Pécs, Pécs, Hungary.

ORCID profiles: Balázs Magyari

Funding

  • National Research, Development, and Innovation Office of Hungary

Cite

APA

Magyari, B., Kittka, B., Goják, I., Kasza, G., Schönfeld, K., Szapáry, L. B., Simon, M., Kiss, R., Bertalan, A., Várady, E., Gyimesi, A., Szokodi, I., & Horváth, I. (2023). Single center experience with the balloon‐expandable Myval transcatheter aortic valve system with the first 100 patients: 30‐day and 1‐year follow‐up. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.30868

Vancouver

Magyari B, Kittka B, Goják I, Kasza G, Schönfeld K, Szapáry LB, et al. Single center experience with the balloon‐expandable Myval transcatheter aortic valve system with the first 100 patients: 30‐day and 1‐year follow‐up. Catheterization and Cardiovascular Interventions. 2023. doi:10.1002/ccd.30868

BibTeX
@article{Magyari2023,
  author = {Magyari, Balázs and Kittka, Bálint and Goják, Ilona and Kasza, Gábor and Schönfeld, Kristóf and Szapáry, László Botond and Simon, Mihály and Kiss, Rudolf and Bertalan, Andrea and Várady, Edit and Gyimesi, András and Szokodi, István and Horváth, Iván},
  title = {Single center experience with the balloon‐expandable Myval transcatheter aortic valve system with the first 100 patients: 30‐day and 1‐year follow‐up},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2023},
  doi = {10.1002/ccd.30868}
}
Start for free →