EUVIMEDDescoperă EUVIMED
Open access · CC BY 4.0

Exploiting Polynomial Chaos Expansion for Rapid Assessment of the Impact of Tissue Property Uncertainties in Low‐Intensity Focused Ultrasound Stimulation

Kemal Sumser, Rob Mestrom, Yunus Emre Tuysuz, Margarethus Marius Paulides

Bioelectromagnetics · 2025

DOI: 10.1002/bem.70004 ↗

At a glance

Journal article in Bioelectromagnetics, 2025. Written by Kemal Sumser and 3 co-authors. Institution: Care & Cure Lab of the Electromagnetics Group (EM4Care+Cure), Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, Th…. Major MEDLINE topics: Nonlinear Dynamics, Ultrasonic Waves, Ultrasonic Therapy. Cited 2 times so far (Crossref); 20 references. The full text is freely available (open access, CC BY 4.0). Funded by: This study was supported by Epilepsiefonds., Epilepsiefonds.

RezumatSubjectsDate bibliograficeCiteRelated studies

Abstract (original language)

Neuromodulation with low‐intensity focused ultrasound (LIFUS) holds significant promise for noninvasive treatment of neurological disorders, but its success relies heavily on accurately targeting specific brain regions. Computational model predictions can be used to optimize LIFUS, but uncertain acoustic tissue properties can affect prediction accuracy. The Monte Carlo method is often used to quantify the impact of uncertainties, but many iterations are generally needed for accurate estimates. We studied a surrogate model based on polynomial chaos expansion (PCE) to quantify the uncertainty in the LIFUS acoustic intensity field caused by tissue acoustic property uncertainties. The PCE approach was benchmarked against Monte Carlo method for LIFUS in three different head models. We also investigated the effect of the number of PCE samples on the accuracy of the surrogate model. Our results show that the PCE surrogate model requires only 20 simulation samples to estimate the mean and standard deviation of the acoustic intensity field with high accuracy compared to 100 samples needed for Monte Carlo method. The root mean squared percentage error (RMSPE) in the mean acoustic intensity field was less than 1.5%, with a maximum error of less than 0.5 W/cm 2 (< 1% of the focus peak intensity in water), while the RMSPE in the standard deviation was less than 9%, with a maximum error of less than 0.3 W/cm 2 . The accuracy of the PCE surrogate model, and the limited number of iterations it requires makes it a promising tool for quantifying the uncertainty in the acoustic intensity field in LIFUS applications.

Subjects

MeSH terms (MEDLINE)

Nonlinear DynamicsUltrasonic WavesUltrasonic TherapyUncertaintyMonte Carlo MethodHumansComputer SimulationBrain

Author keywords

Monte Carlo methodfocused ultrasoundneuromodulationpolynomial chaos expansionultrasound

Authors and institutions

  • Care & Cure Lab of the Electromagnetics Group (EM4Care+Cure), Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands.
  • Department of Electrical and Electronics Engineering, Middle East Technical University, Ankara, Turkey.
  • Department of Radiotherapy, Erasmus University Medical Center Cancer Institute, Rotterdam, The Netherlands.

Funding

  • This study was supported by Epilepsiefonds.
  • Epilepsiefonds

Cite

APA

Sumser, K., Mestrom, R., Tuysuz, Y. E., & Paulides, M. M. (2025). Exploiting Polynomial Chaos Expansion for Rapid Assessment of the Impact of Tissue Property Uncertainties in Low‐Intensity Focused Ultrasound Stimulation. Bioelectromagnetics. https://doi.org/10.1002/bem.70004

Vancouver

Sumser K, Mestrom R, Tuysuz YE, Paulides MM. Exploiting Polynomial Chaos Expansion for Rapid Assessment of the Impact of Tissue Property Uncertainties in Low‐Intensity Focused Ultrasound Stimulation. Bioelectromagnetics. 2025. doi:10.1002/bem.70004

BibTeX
@article{Sumser2025,
  author = {Sumser, Kemal and Mestrom, Rob and Tuysuz, Yunus Emre and Paulides, Margarethus Marius},
  title = {Exploiting Polynomial Chaos Expansion for Rapid Assessment of the Impact of Tissue Property Uncertainties in Low‐Intensity Focused Ultrasound Stimulation},
  journal = {Bioelectromagnetics},
  year = {2025},
  doi = {10.1002/bem.70004}
}
Start for free →