Characterization of weakly nonlinear effects in relationship to transducer parameters in focused ultrasound therapy

Peng Xu, Hao Wu, Guofeng Shen

Medical Physics · 2024

DOI: 10.1002/mp.17270 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Peng Xu und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, People's Republic of China. Hauptthemen laut MEDLINE: Transducers, Nonlinear Dynamics. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 62 Literaturangaben. Gefördert von: National Key Research and Development Program of the Ministry of Science and Technology, National Natural Science Foundation of China, China Scholarship Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Focused ultrasound therapy has been widely used for the treatment of various diseases, employing different types of transducers. The focused ultrasound pressure fields inevitably exhibit nonlinear effects, which can influence the ablation region. However, the nonlinear effects exhibit noticeable variations across different applications. The characterization of the nonlinear pressure fields of ultrasound is important for the effective implementation of focused ultrasound therapy.

Purpose

The traditional angular spectrum method (ASM) was extended to accurately and efficiently simulate the propagation of weakly nonlinear ultrasound in heterogeneous mediums of clinical model. The nonlinear effects were further analyzed in relationship to the transducer parameters that are different in various applications.

Methods

The pressure fields were simulated using the extended ASM, incorporating calculations for phase aberration in the frequency domain and magnitude compensation in the spatial domain to account for heterogeneous acoustic impedance mismatch. Validation was performed by comparison to k‐Wave simulation results using two simplified clinical models, an abdominal soft tissue and a transcranial skull model. The nonlinear effects were then analyzed in relation to the transducer parameters of f‐number and effective source area based on the same acoustic output power. The analysis of nonlinear effects was conducted under both homogeneous medium and the clinical models.

Results

The simulation results demonstrated a maximum error of 3.93% in the calculated harmonic pressure of the abdominal model, and a maximum error of 4.89% within the transcranial model when comparing the extended ASM simulation results to those obtained from k‐Wave simulations. The characterization of the nonlinear effects reveals a strong correlation with the transducer parameters. Specifically, the results indicate that the nonlinear effects intensify with an increase in the effective source area and f‐number, under the same acoustic output power of the transducer. However, the clinical model also showed an influence on the nonlinear effects in relation to the f‐number.

Conclusion

The extended ASM was demonstrated as an accurate and efficient simulation tool, and the simulation results provide a reference for evaluating the intensity of nonlinear effects in various transducer designs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TransducersNonlinear DynamicsHigh-Intensity Focused Ultrasound AblationUltrasonic TherapyPressureHumansSkullAbdomenComputer Simulation

Stichwörter der Autor:innen

focused ultrasoundnonlinear effectspressure field simulation

Autor:innen und Institutionen

  • School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, People's Republic of China.
  • Shanghai Shende Green Medical Era Healthcare Technology Co., Ltd, Shanghai, People's Republic of China.

Förderung

  • National Key Research and Development Program of the Ministry of Science and Technology
  • National Natural Science Foundation of China
  • China Scholarship Council
  • Shanghai Technical leader

Zitieren

APA

Xu, P., Wu, H., & Shen, G. (2024). Characterization of weakly nonlinear effects in relationship to transducer parameters in focused ultrasound therapy. Medical Physics. https://doi.org/10.1002/mp.17270

Vancouver

Xu P, Wu H, Shen G. Characterization of weakly nonlinear effects in relationship to transducer parameters in focused ultrasound therapy. Medical Physics. 2024. doi:10.1002/mp.17270

BibTeX
@article{Xu2024,
  author = {Xu, Peng and Wu, Hao and Shen, Guofeng},
  title = {Characterization of weakly nonlinear effects in relationship to transducer parameters in focused ultrasound therapy},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17270}
}
Kostenlos starten →