EUVIMEDObjevte EUVIMED
Open access · CC BY 4.0

Selection of DNA Aptamers Against Parathyroid Hormone for Electrochemical Impedimetric Biosensor System Development

Reza Didarian, Saharnaz Bargh, Almina Gülerman, Veli Cengiz Özalp, Özcan Erel, Nimet Yildirim-Tirgil

Biotechnology and Applied Biochemistry · 2025-10 · 72 (5) · 1351-1362

DOI: 10.1002/bab.2745 ↗

At a glance

Journal article in Biotechnology and Applied Biochemistry, published online on 2025-03-19. Written by Reza Didarian and 5 co-authors. Institution: Department of Biomedical Engineering, Ankara Yıldırım Beyazıt University, Ankara, Türkiye. Major MEDLINE topics: Aptamers, Nucleotide, Parathyroid Hormone, Biosensing Techniques, SELEX Aptamer Technique. Cited 11 times so far (Crossref); 42 references. The full text is freely available (open access, CC BY 4.0). Funded by: Türkiye Bilimsel ve Teknolojik Araştırma Kurumu.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

This work presents the pioneering development of an aptamer‐based electrochemical biosensor for real‐time monitoring of parathyroid hormone (PTH) levels, with a focus on intraoperative assessment during parathyroid surgery. It introduces, for the first time, the selection and characterization of aptamers targeting distinct segments of the PTH peptide. The study demonstrates the feasibility and efficacy of the biosensing platform through a precisely designed experimental framework, including SELEX‐based aptamer selection, aptamer–peptide interaction analysis, and biosensor fabrication. The SELEX process yields aptamers with notable binding affinities to different fragments of PTH, with the PTH (53–84) aptamer showing particularly sensitive binding to the hormone's C terminus, allowing for precise PTH analysis. Electrochemical characterization reveals significant changes in electrochemical impedance spectroscopy (EIS) signals upon exposure to varying PTH concentrations, highlighting the sensor's sensitivity and selectivity. The increase in charge transfer resistance (Rct) values with rising PTH concentrations underscores the biosensor's capability to detect PTH‐induced structural changes, validating its potential for accurate measurement. The biosensor shows remarkable selectivity in the presence of common interferents in serum samples, ensuring precise PTH detection. Stability assessments over a 45‐day storage period demonstrate the biosensor's robustness and long‐term reliability, affirming its practical suitability. In summary, the developed aptamer‐based biosensor represents a promising tool for sensitive and selective PTH detection, with potential applications in biomedical research and clinical diagnostics, particularly for intraoperative PTH analysis during parathyroidectomy. Continued research and optimization efforts hold promise for enhancing its performance and expanding its utility in diverse healthcare settings.

Subjects

MeSH terms (MEDLINE)

Aptamers, NucleotideParathyroid HormoneBiosensing TechniquesSELEX Aptamer TechniqueElectrochemical TechniquesHumansDielectric Spectroscopy

Author keywords

SELEXaptamerelectrochemical detectionelectrochemical impedance spectroscopy (EIS)impedimetric biosensorparathyroid hormone (PTH)

Substances

Aptamers, NucleotideParathyroid Hormone

Authors and institutions

  • Department of Biomedical Engineering, Ankara Yıldırım Beyazıt University, Ankara, Türkiye.
  • Department of Radiopharmacy, Faculty of Pharmacy, Hacettepe University, Ankara, Türkiye.
  • Department of Metallurgy and Materials Engineering, Ankara Yıldırım Beyazıt University, Ankara, Türkiye.
  • Department of Medical Biology, School of Medicine, Atılım University, Ankara, Türkiye.
  • Department of Biovchemistry, Faculty of Medicine, Ankara Yıldırım Beyazıt University, Ankara, Türkiye.

Funding

  • Türkiye Bilimsel ve Teknolojik Araştırma Kurumu

Cite

APA

Didarian, R., Bargh, S., Gülerman, A., Özalp, V. C., Erel, Ö., & Yildirim-Tirgil, N. (2025). Selection of DNA Aptamers Against Parathyroid Hormone for Electrochemical Impedimetric Biosensor System Development. Biotechnology and Applied Biochemistry, 72(5), 1351-1362. https://doi.org/10.1002/bab.2745

Vancouver

Didarian R, Bargh S, Gülerman A, Özalp VC, Erel Ö, Yildirim-Tirgil N. Selection of DNA Aptamers Against Parathyroid Hormone for Electrochemical Impedimetric Biosensor System Development. Biotechnology and Applied Biochemistry. 2025;72(5):1351-1362. doi:10.1002/bab.2745

BibTeX
@article{Didarian2025,
  author = {Didarian, Reza and Bargh, Saharnaz and Gülerman, Almina and Özalp, Veli Cengiz and Erel, Özcan and Yildirim-Tirgil, Nimet},
  title = {Selection of DNA Aptamers Against Parathyroid Hormone for Electrochemical Impedimetric Biosensor System Development},
  journal = {Biotechnology and Applied Biochemistry},
  year = {2025},
  volume = {72},
  number = {5},
  pages = {1351--1362},
  doi = {10.1002/bab.2745}
}