Journal article in Advanced Science, 2026. Written by Karteekeya Sastry and 7 co-authors. Institution: Caltech Optical Imaging Laboratory, Andrew and Peggy Cherng Department of Medical Engineering, Department of Electrical Engineering, California Insti…. Major MEDLINE topics: Photoacoustic Techniques, Oximetry, Oxygen, Radial Artery. Cited 1 times so far (Crossref); 64 references. The full text is freely available (open access, CC BY 4.0). Funded by: NIH HHS, NIDA NIH HHS, NCI NIH HHS.

AbstraktSubjectsBibliograafilised andmedCiteRelated studies

Abstract (original language)

Photoacoustic tomography is uniquely capable of high‐resolution deep‐tissue blood‐oxygenation (sO 2 ) imaging (oximetry) due to its optical absorption contrast. However, wavelength‐dependent optical fluence changes within tissue, i.e., spectral coloring, have impeded the development of photoacoustic oximetry. We present the arterial prior method (APM+; + denotes intravascular fluence correction), which leverages the high arterial sO 2 to locally calibrate the optical fluence within tissue to circumvent spectral coloring and reliably estimate the sO 2 near the artery. In phantom experiments with ex vivo animal tissue, APM+ resulted in a median estimation error of 2.9% compared to 9.8% from the traditional linear unmixing method (LUM). In human imaging experiments of the radial artery‐vein pair in eight healthy adult volunteers, the estimated venous sO 2 s from APM+ (median: 72.3%, interquartile range/IQR: 8.9%) were concentrated around the typical 60%–80% range in healthy individuals, whereas those from LUM (median: 75.2%, IQR: 34.4%) varied widely. When imaging the wrists of the eight subjects through ex vivo animal tissue of thicknesses up to 1.5 cm, APM+ provided more consistent estimates than LUM, indicating its robustness with depth.

Subjects

MeSH terms (MEDLINE)

Photoacoustic TechniquesOximetryOxygenRadial ArteryHumansAnimalsAdultPhantoms, ImagingMaleFemale

Author keywords

arterial bloodbiomedical engineeringoxygen saturationphotoacoustic tomography

Substances

Oxygen

Authors and institutions

  • Caltech Optical Imaging Laboratory, Andrew and Peggy Cherng Department of Medical Engineering, Department of Electrical Engineering, California Institute of Technology, Pasadena, California, USA.
  • College of Biomedical Engineering and Instrument Science, Zhejiang University, Hangzhou, China.
  • Department of Biosciences and Bioengineering, Indian Institute of Technology Jammu, Jammu, India.

ORCID profiles: Karteekeya Sastry, Junhao Zhu, Joshua Olick-Gibson, Li Lin, Lei Li, Jigmi Basumatary

Funding

  • NIH HHS
  • NIDA NIH HHS
  • NCI NIH HHS
  • NIBIB NIH HHS
  • National Institutes of Health

Cite

APA

Sastry, K., Zhu, J., Olick-Gibson, J., Lin, L., Li, L., Basumatary, J., Zhang, Y., & Wang, L. V. (2026). Arterial Blood‐Mediated Deep‐Tissue Photoacoustic Oximetry. Advanced Science. https://doi.org/10.1002/advs.76366

Vancouver

Sastry K, Zhu J, Olick-Gibson J, Lin L, Li L, Basumatary J, et al. Arterial Blood‐Mediated Deep‐Tissue Photoacoustic Oximetry. Advanced Science. 2026. doi:10.1002/advs.76366

BibTeX
@article{Sastry2026,
  author = {Sastry, Karteekeya and Zhu, Junhao and Olick-Gibson, Joshua and Lin, Li and Li, Lei and Basumatary, Jigmi and Zhang, Yang and Wang, Lihong V},
  title = {Arterial Blood‐Mediated Deep‐Tissue Photoacoustic Oximetry},
  journal = {Advanced Science},
  year = {2026},
  doi = {10.1002/advs.76366}
}