EUVIMEDEUVIMED entdecken
Crossref

Clinical Validation of a Wearable Piezoelectric Blood‐Pressure Sensor for Continuous Health Monitoring

Seongwook Min, Dong Hyun Kim, Daniel J Joe, Byung Woo Kim, Young Hoon Jung, Jae Hee Lee, Bo-Yeon Lee, Il Doh, Jaehun An, Young-Nam Youn, Boyoung Joung, Chang D Yoo und 2 weitere

Advanced Materials · 2023-06 · 35 (26)

DOI: 10.1002/adma.202301627 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Advanced Materials, online erschienen am 11.05.2023. Verfasst von Seongwook Min und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Materials Science and Engineering, Korea Advanced Institute of Science and Technology (KAIST), 291 Daehak-ro, Yuseong-gu, Daejeon, 3414…. Hauptthemen laut MEDLINE: Arterial Pressure, Wearable Electronic Devices. Die Arbeit wurde bisher 282-mal zitiert (Crossref) und nennt 52 Literaturangaben. Gefördert von: Korean Government, National Research Foundation of Korea, Korea government.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Wearable blood‐pressure sensors have recently attracted attention as healthcare devices for continuous non‐invasive arterial pressure (CNAP) monitoring. However, the accuracy of wearable blood‐pressure (BP) monitoring devices has been controversial due to the low signal quality of sensors, the absence of an accurate transfer function to convert the sensor signals into BP values, and the lack of clinical validation regarding measurement precision. Here, a wearable piezoelectric blood‐pressure sensor (WPBPS) is reported, which achieves a high normalized sensitivity (0.062 kPa −1 ), and fast response time (23 ms) for CNAP monitoring. The transfer function of a linear regression model is designed, offering a simple solution to convert the flexible piezoelectric sensor signals into BP values. In order to verify the measurement accuracy of WPBPS, clinical trials are performed on 35 subjects aged from 20 to 80 s after screening. The mean difference between the WPBPS and a commercial sphygmomanometer of 175 BP data pairs is −0.89 ± 6.19 and −0.32 ± 5.28 mmHg for systolic blood pressure (SBP) and diastolic blood pressure (DBP), respectively. By building a WPBPS‐embedded wristwatch, the potentially promising use of a convenient, portable, continuous BP monitoring system for cardiovascular disease diagnosis is demonstrated.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arterial PressureWearable Electronic DevicesHumansBlood PressureBlood Pressure DeterminationBlood Pressure Monitors

Stichwörter der Autor:innen

clinical validationcontinuous and non-invasive arterial pressure (CNAP) monitoringflexible piezoelectric sensorspulse wave analysiswearable piezoelectric blood-pressure sensors

Autor:innen und Institutionen

  • Department of Materials Science and Engineering, Korea Advanced Institute of Science and Technology (KAIST), 291 Daehak-ro, Yuseong-gu, Daejeon, 34141, Republic of Korea.
  • Safety Measurement Institute, Korea Research Institute of Standards and Science (KRISS), 267 Gajeong-ro, Yuseong-gu, Daejeon, 34113, Republic of Korea.
  • Department of Nature-Inspired System and Application, Korea Institute of Machinery and Materials (KIMM), 156 Gajeongbuk-ro, Yuseong-gu, Daejeon, 34103, Republic of Korea.
  • Division of Cardiovascular Surgery, Severance Cardiovascular Hospital, Yonsei University, College of Medicine, Yonsei University Health System, 50 Yonsei-ro, Seodaemun-gu, Seoul, 03722, Republic of Korea.
  • Division of Cardiology, Severance Cardiovascular Hospital, Yonsei University College of Medicine, 50 Yonsei-ro, Seodaemun-gu, Seoul, 03722, Republic of Korea.
  • Department of Electrical Engineering, Korea Advanced Institute of Science and Technology (KAIST), 291 Daehak-ro, Yuseong-gu, Daejeon, 34141, Republic of Korea.

ORCID-Profile: Keon Jae Lee

Förderung

  • Korean Government
  • National Research Foundation of Korea
  • Korea government
  • Development of Measurement Standards and Technology
  • Medical Convergence funded by Korea Research Institute of Standards and Science

Zitieren

APA

Min, S., Kim, D. H., Joe, D. J., Kim, B. W., Jung, Y. H., Lee, J. H., Lee, B. Y., Doh, I., An, J., Youn, Y. N., Joung, B., Yoo, C. D., Ahn, H. S., & Lee, K. J. (2023). Clinical Validation of a Wearable Piezoelectric Blood‐Pressure Sensor for Continuous Health Monitoring. Advanced Materials, 35(26). https://doi.org/10.1002/adma.202301627

Vancouver

Min S, Kim DH, Joe DJ, Kim BW, Jung YH, Lee JH, et al. Clinical Validation of a Wearable Piezoelectric Blood‐Pressure Sensor for Continuous Health Monitoring. Advanced Materials. 2023;35(26). doi:10.1002/adma.202301627

BibTeX
@article{Min2023,
  author = {Min, Seongwook and Kim, Dong Hyun and Joe, Daniel J and Kim, Byung Woo and Jung, Young Hoon and Lee, Jae Hee and Lee, Bo-Yeon and Doh, Il and An, Jaehun and Youn, Young-Nam and Joung, Boyoung and Yoo, Chang D and Ahn, Hyo-Suk and Lee, Keon Jae},
  title = {Clinical Validation of a Wearable Piezoelectric Blood‐Pressure Sensor for Continuous Health Monitoring},
  journal = {Advanced Materials},
  year = {2023},
  volume = {35},
  number = {26},
  doi = {10.1002/adma.202301627}
}
Kostenlos starten →