Accuracy and Safety of a Continuous Noninvasive Blood Pressure Monitor in Neonates
Neonatology · 2025
DOI: 10.1159/000546187 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Neonatology, 2025. Verfasst von R Brandon Hunter und 10 weiteren Autor:innen. Beteiligte Einrichtung: Texas Children's Hospital, Houston, Texas, USA. Hauptthemen laut MEDLINE: Blood Pressure Monitors, Blood Pressure Determination, Blood Pressure. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: FDA HHS, NICHD NIH HHS.
Abstract (Originalsprache)
Introduction
Accurate and continuous blood pressure (BP) monitoring in neonates is crucial in the intensive care unit. Invasive arterial lines (IALs), oscillometric cuffs, and current noninvasive continuous BP monitoring devices have significant limitations. The Boppli® device is a novel, continuous, noninvasive BP device that requires no calibration, designed for neonates.
Methods
This prospective, multicenter study evaluated the performance, usability, and safety of the Boppli device in neonates ±5 mm Hg. Various subgroups had SDs >8 mm Hg attributed to low sample sizes. The device received high usability scores from clinicians and parents. No serious adverse events were reported.
Conclusion
The Boppli device is a promising alternative for continuous noninvasive BP monitoring in neonates, offering good accuracy and usability. The device, which received 510(k) clearance in September 2023, was well received by clinicians and parents, with a low-risk profile.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Texas Children's Hospital, Houston, Texas, USA.
- Baylor College of Medicine, Houston, Texas, USA.
- Children's National Hospital, Washington, DC, USA.
- McGovern Medical School, UT Health, Houston, Texas, USA.
- Massachusetts General Hospital, Boston, Massachusetts, USA.
- Tufts Medical Center, Boston, Massachusetts, USA.
Förderung
- FDA HHS
- NICHD NIH HHS
Zitieren
APA
Hunter, R. B., Saruwatari, M. S., Bliss, N. D., Kim, S., Turcu, R., Krbec, B. A., Yusuf, K., Rao, A., Quan, X., Soghier, L., & Boppli Research Group (2025). Accuracy and Safety of a Continuous Noninvasive Blood Pressure Monitor in Neonates. Neonatology. https://doi.org/10.1159/000546187
Vancouver
Hunter RB, Saruwatari MS, Bliss ND, Kim S, Turcu R, Krbec BA, et al. Accuracy and Safety of a Continuous Noninvasive Blood Pressure Monitor in Neonates. Neonatology. 2025. doi:10.1159/000546187