Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devices
Journal of Applied Clinical Medical Physics · 2026-12 · 27 (12)
DOI: 10.1002/acm2.70913 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 17.11.2026. Verfasst von Inbum Lee und 9 weiteren Autor:innen. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Routine quality assurance (QA) of linear accelerators and electronic portal imaging devices (EPIDs) often requires separate devices and acquisition procedures for geometric, image‐quality, radiation‐field, multileaf collimator (MLC), and dosimetric checks. A multifunctional phantom could consolidate these assessments within a single QA workflow.
Purpose
To evaluate the week‐to‐week reproducibility of measurements obtained with MZ Slab, a multifunctional megavoltage EPID phantom, on three linear accelerators with different EPID configurations.
Methods
Measurements were performed weekly for four weeks on TrueBeam STx, Halcyon, and Harmony Pro. Each session comprised three acquisition sets with phantom removal and repositioning between sets. The phantom provided measurements of geometric accuracy, image quality, material‐response isotropy, and radiation‐field size, together with qualitative MLC light‐field checks. Weekly values were summarized as mean and within‐session standard deviation, and changes were assessed relative to week 1. Field sizes were compared with open‐field EPID images and IC Profiler measurements. Dose‐calibrated portal images acquired on TrueBeam were compared with the week 1 image using mean response change and gamma passing rates.
Results
Central‐marker offset ranged from 0.24 to 0.78 mm across units, with a maximum change from week 1 of 0.29 mm. Marker‐distance deviations changed by no more than 0.03 mm. The maximum changes in mean relative contrast transfer function and low‐contrast contrast‐to‐noise ratio were 0.011 (2.6%) and 1.81 (5.0%), respectively, relative to the week 1 baseline. Uniformity deviation and normalized noise changed by up to 0.50 and 0.17 percentage points, and material‐specific normalized response by up to 0.014. Field sizes changed by up to 0.56 mm and agreed with open‐field EPID measurements within 0.3 mm. IC Profiler values were 0.2–0.9 mm larger than MZ Slab values. MLC leaf ends remained within the printed ± 1 mm tolerance. At week 4, mean portal‐image response had increased by 0.77%, with a gamma passing rate of 99.65% at 1%/1 mm.
Conclusions
MZ Slab supported geometric, image‐quality, radiation‐field, qualitative MLC, and portal‐dosimetry assessments within a single‐phantom QA workflow. Measurements on three linacs characterized within‐session reproducibility and week‐to‐week consistency over four weeks. Longer observation and deliberate perturbation testing are required to establish action limits and fault‐detection sensitivity.
Zitieren
APA
Lee, I., Choi, M., Huh, Y., Yoon, E., Cheon, B., Kwon, N. H., Jang, Y. J., Choi, C. H., Kim, J., & Park, S. (2026). Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devices. Journal of Applied Clinical Medical Physics, 27(12). https://doi.org/10.1002/acm2.70913
Vancouver
Lee I, Choi M, Huh Y, Yoon E, Cheon B, Kwon NH, et al. Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devices. Journal of Applied Clinical Medical Physics. 2026;27(12). doi:10.1002/acm2.70913
BibTeX
@article{Lee2026,
author = {Lee, Inbum and Choi, Minjae and Huh, Yoonsuk and Yoon, Euntaek and Cheon, Bo‐Wi and Kwon, Na Hye and Jang, Young Jae and Choi, Chang Heon and Kim, Jung‐in and Park, So‐Yeon},
title = {Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devices},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {12},
doi = {10.1002/acm2.70913}
}
EUVIMED