Monte Carlo simulations of geometric deformation of Harrison–Anderson–Mick applicators used in intraoperative radiation therapy
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70386 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Benjamin Insley und 9 weiteren Autor:innen. Beteiligte Einrichtung: Empyrean Medical Systems, Inc, Boca Raton, Florida, USA. Hauptthemen laut MEDLINE: Monte Carlo Method, Radiotherapy Planning, Computer-Assisted, Brachytherapy, Phantoms, Imaging. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The Harrison–Anderson–Mick (HAM) applicator is a high‐dose‐rate intraoperative radiotherapy (HDR‐IORT) applicator used to position Ir‐192 brachytherapy sources over surgically‐accessed tumor volumes or post‐resection tumor beds. Because of the lack of a 3D imaging system, dwell times are optimized pre‐surgery by a TG‐43–based treatment planning system (TPS) that assumes a perfectly flat applicator surface surrounded by an infinite water phantom. These assumed conditions are disparate from typical treatment conditions, especially in the pelvic regions, which often involve uneven patient surfaces and superficial irradiations with little to no backscatter material.
Purpose
Develop and validate a Monte Carlo (MC) model of HDR‐IORT treatment with a HAM applicator and use this validated model to quantify inaccuracies in the dose calculations due to the simplified conditions assumed in the planning process.
Methods
The Oncentra Brachy TPS was used to optimize dwell times and calculate the delivered relative dose distribution for a 6‐catheter, 8‐cm HAM applicator with 0.5‐cm dwell steps and 1.0‐cm catheter spacing. A TOPAS (v.3.9) MC model of this treatment was then developed and validated against the TPS dose calculations. Once validated, the TOPAS applicator model was modified to calculate the difference in doses due to changes in backscatter conditions, the incorporation of the actual applicator materials, and curvature of the applicator. Dose distributions were characterized using dose to the prescription point, percent depth doses, and two‐dimensional isodose curves.
Results
Validation between the MC and TPS calculations was successful within 1.5% over a depth of 5.0 cm in water. Negligible dose differences were calculated when assuming the applicator is made completely of water versus modeling all the individual applicator components. Conversely, semi‐infinite phantom geometry had more than 5% loss in dose to the prescription point due to the absence of backscatter. Most severely, curvature radii in the range of 5.4 cm (shallow curvature) to 0.9 cm (steep curvature) had dose differences of 5%–25%, regardless of whether the curvature was along the catheter length or transverse to it.
Conclusions
This study quantified the changes in dose due to material and geometric differences that are currently not accounted for by the TPS. While not accounting for the material of the applicator contributes to negligibly, the presence of backscattering material can contribute up to 5%. The radius of the bending of the applicator was found to potentially have the largest impact on the dosimetry of the central plane, with deviations up to 25%.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Empyrean Medical Systems, Inc, Boca Raton, Florida, USA.
- University of Miami, Miami, Florida, USA.
- Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- The University of Texas MD Anderson Cancer Center UTHealth Houston Graduate School of Biomedical Sciences, Houston, Texas, USA.
- Accuray Inc., Sunnyvale, California, USA.
ORCID-Profile: Brett Bocian, Patrick James Jensen, Rachael M Martin-Paulpeter, Joshua Scott Niedzielski, Luis Augusto Perles
Zitieren
APA
Insley, B., Bocian, B., Beddar, S., Reiazi, R., Jensen, P. J., Martin-Paulpeter, R. M., Niedzielski, J. S., Flint, D., Sawakuchi, G. O., & Perles, L. A. (2025). Monte Carlo simulations of geometric deformation of Harrison–Anderson–Mick applicators used in intraoperative radiation therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70386
Vancouver
Insley B, Bocian B, Beddar S, Reiazi R, Jensen PJ, Martin-Paulpeter RM, et al. Monte Carlo simulations of geometric deformation of Harrison–Anderson–Mick applicators used in intraoperative radiation therapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70386
BibTeX
@article{Insley2025,
author = {Insley, Benjamin and Bocian, Brett and Beddar, Sam and Reiazi, Reza and Jensen, Patrick James and Martin-Paulpeter, Rachael M and Niedzielski, Joshua Scott and Flint, David and Sawakuchi, Gabriel Oliveira and Perles, Luis Augusto},
title = {Monte Carlo simulations of geometric deformation of Harrison–Anderson–Mick applicators used in intraoperative radiation therapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70386}
}
EUVIMED