EUVIMEDEUVIMED entdecken
ÜbersichtsarbeitOpen Access · CC BY 4.0

Vision and implementation for the first U.S. carbon ion therapy center part II: Design philosophy and clinical implementation

Xiaoying Liang iD, Chris J Beltran

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70866 ↗

Kurz gefasst

Übersichtsarbeit in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Xiaoying Liang und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Heavy Ion Radiotherapy, Neoplasms, Quality Assurance, Health Care. Die Arbeit nennt 65 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The anticipated clinical demand for carbon ion radiotherapy (CIRT) in the United States is expected to include a substantial proportion of clinically complex disease sites that require accurate, highly conformal, and adaptive treatment. These clinical needs, together with the unique physical and biological characteristics of carbon ion beams, form the foundation of the design philosophy for the Mayo Clinic Florida (MCF) CIRT program. The program remains under active development, with clinical deployment planned for 2028. Building upon the clinical rationale and anticipated U.S. patient population discussed in Part I of this two‐part review, this Vision Article describes how MCF intends to develop an integrated CIRT treatment system. The program is guided by two overarching principles: fast, accurate, and safe beam delivery and accurate treatment planning and treatment delivery, supported by an integrated patient‐specific quality assurance framework. Key elements include high‐dose‐rate dose‐driven continuous scanning with advanced beam monitoring and irradiation control, delivery‐aware treatment planning, the Monte Carlo‐based BioPhys computational platform integrating physical and biological dose modeling, a hierarchical adaptive workflow, and patient‐specific quality assurance spanning pretreatment, on‐treatment, and post‐treatment verification. This framework provides a technical roadmap for the integrated clinical implementation of CIRT at MCF.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedHeavy Ion RadiotherapyNeoplasmsQuality Assurance, Health CareHumansRadiotherapy DosageUnited StatesMonte Carlo Method

Autor:innen und Institutionen

  • Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA.

ORCID-Profile: Xiaoying Liang

Zitieren

APA

Liang, X., & Beltran, C. J. (2026). Vision and implementation for the first U.S. carbon ion therapy center part II: Design philosophy and clinical implementation. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70866

Vancouver

Liang X, Beltran CJ. Vision and implementation for the first U.S. carbon ion therapy center part II: Design philosophy and clinical implementation. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70866

BibTeX
@article{Liang2026,
  author = {Liang, Xiaoying and Beltran, Chris J},
  title = {Vision and implementation for the first U.S. carbon ion therapy center part II: Design philosophy and clinical implementation},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70866}
}
Kostenlos starten →