EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Density override‐based VMAT optimization to improve robustness at tissue‐air interfaces in head and neck radiotherapy

Sagarika Jain, Ashlee Ewing, Michael Weldon, Allison Palmiero, Daniel Christ, Emile Gogineni, Simeng Zhu, Sung Jun Ma, John C Grecula, Darrion L Mitchell, Sujith Baliga, Dukagjin M Blakaj und 1 weitere

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70737 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Sagarika Jain und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, James Cancer Hospital, The Ohio State University, Columbus, Ohio, USA. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiotherapy Planning, Computer-Assisted, Head and Neck Neoplasms, Organs at Risk. Die Arbeit nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Volumetric Modulated Arc Therapy (VMAT) is integral to head and neck (HN) radiation therapy; however, plan robustness may be compromised at air‐tissue interfaces. Highly modulated fluence directed at these areas can produce significant unexpected high‐dose regions in the event of minor variations in tissue density.

Purpose

This study characterizes an instability in VMAT planning for HN radiation therapy (RT) occurring at the interface between target volume (e.g. mucosal primary tumor) and internal air spaces (e.g. pharyngeal lumen, sinuses, nasal cavity, etc.). A density‐override based planning technique is presented and validated to mitigate this instability.

Methods

Treatment plans for fifteen HN patients receiving VMAT RT at our institution between 2023 and 2024 were retrospectively analyzed. All patients received 60–69.96 Gy in 30–33 daily fractions using 6 MV photons. To model swelling or tumor growth near the air–tumor interface, verification plans were generated by overriding internal air within 3 mm of the planning target volume (PTV) to soft tissue density (HU = 0). A robustness‐enhancing technique was evaluated, wherein optimization was performed with luminal air overridden to HU = –300, followed by recalculation and renormalization using the original CT HU values. Robustness was then reassessed after overriding luminal air to HU = 0.

Results

Baseline plans demonstrated controlled hotspot magnitudes within the BODY, with a median D0.1cc of 109.0% [Interquartile range: 107.8–110.4]. Simulated tissue‐air interface changes produced substantially elevated hotspots in verification plans, increasing to a median D0.1cc of 115.6% [111.9–126.9], corresponding to a median paired increase of 9.0% (Wilcoxon p = 1.22 × 10 − 4 ). Robust plans generated using the proposed density‐override optimization maintained comparable baseline dosimetry (median D0.1cc 108.1% [107.4–108.7]) while demonstrating stable hotspot behavior under simulated density perturbations (median D0.1cc 108.1% [107.1–108.6], p = 0.18). Target coverage remained preserved, with robust verification plan D95 values ranging from 99.7% to 100.5% ( p = 0.26).

Conclusion

A density‐override optimization approach mitigates hotspot formation at air–tumor interfaces, producing VMAT plans that remain dosimetrically stable despite variation at internal air‐tumor interfaces.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedHead and Neck NeoplasmsOrgans at RiskHumansRadiotherapy DosageRetrospective StudiesAlgorithmsAirPrognosis

Stichwörter der Autor:innen

HNVMATdensity‐override

Autor:innen und Institutionen

  • Department of Radiation Oncology, James Cancer Hospital, The Ohio State University, Columbus, Ohio, USA.

Zitieren

APA

Jain, S., Ewing, A., Weldon, M., Palmiero, A., Christ, D., Gogineni, E., Zhu, S., Ma, S. J., Grecula, J. C., Mitchell, D. L., Baliga, S., Blakaj, D. M., & Konieczkowski, D. J. (2026). Density override‐based VMAT optimization to improve robustness at tissue‐air interfaces in head and neck radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70737

Vancouver

Jain S, Ewing A, Weldon M, Palmiero A, Christ D, Gogineni E, et al. Density override‐based VMAT optimization to improve robustness at tissue‐air interfaces in head and neck radiotherapy. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70737

BibTeX
@article{Jain2026,
  author = {Jain, Sagarika and Ewing, Ashlee and Weldon, Michael and Palmiero, Allison and Christ, Daniel and Gogineni, Emile and Zhu, Simeng and Ma, Sung Jun and Grecula, John C and Mitchell, Darrion L and Baliga, Sujith and Blakaj, Dukagjin M and Konieczkowski, David J},
  title = {Density override‐based VMAT optimization to improve robustness at tissue‐air interfaces in head and neck radiotherapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70737}
}
Kostenlos starten →