Evaluating automatically generated normal tissue contours for safe use in head and neck and cervical cancer treatment planning
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14338 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Raphael Douglas und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Head and Neck Neoplasms, Radiotherapy, Intensity-Modulated, Organs at Risk. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Volumetric‐modulated arc therapy (VMAT) is a widely accepted treatment method for head and neck (HN) and cervical cancers; however, creating contours and plan optimization for VMAT plans is a time‐consuming process. Our group has created an automated treatment planning tool, the Radiation Planning Assistant (RPA), that uses deep learning models to generate organs at risk (OARs), planning structures and automates plan optimization. This study quantitatively evaluates the quality of contours generated by the RPA tool.
Methods
For patients with HN (54) and cervical (39) cancers, we retrospectively generated autoplans using the RPA. Autoplans were generated using deep‐learning and RapidPlan models developed in‐house. The autoplans were, then, applied to the original, physician‐drawn contours, which were used as a ground truth (GT) to compare with the autocontours (RPA). Using a “two one‐sided tests” (TOST) procedure, we evaluated whether the autocontour normal tissue dose was equivalent to that of the ground truth by a margin, δ , that we determined based on clinical judgement. We also calculated the number of plans that met established clinically accepted dosimetric criteria.
Results
For HN plans, 91.8% and 91.7% of structures met dosimetric criteria for automatic and manual contours, respectively; for cervical plans, 95.6% and 95.7% of structures met dosimetric criteria for automatic and manual contours, respectively. Autocontours were equivalent to the ground truth for 71% and 75% of common DVH metrics for the HN and cervix, respectively.
Conclusions
This study shows that dosimetrically equivalent normal tissue contours can be created for HN and cervical cancers using deep learning techniques. In general, differences between the contours did not affect the passing or failing of clinical dose tolerances.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Radiation Oncology, Stanford University, Stanford, California, USA.
ORCID-Profile: Beth M Beadle
Zitieren
APA
Douglas, R., Olanrewaju, A., Mumme, R., Zhang, L., Beadle, B. M., & Court, L. E. (2024). Evaluating automatically generated normal tissue contours for safe use in head and neck and cervical cancer treatment planning. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14338
Vancouver
Douglas R, Olanrewaju A, Mumme R, Zhang L, Beadle BM, Court LE. Evaluating automatically generated normal tissue contours for safe use in head and neck and cervical cancer treatment planning. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14338
BibTeX
@article{Douglas2024,
author = {Douglas, Raphael and Olanrewaju, Adenike and Mumme, Raymond and Zhang, Lifei and Beadle, Beth M and Court, Laurence Edward},
title = {Evaluating automatically generated normal tissue contours for safe use in head and neck and cervical cancer treatment planning},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14338}
}
EUVIMED