EUVIMEDEUVIMED entdecken
Multizentrische StudieOpen Access · CC BY 4.0

Application of an updated severity‐based FMEA approach to Radiation Therapy QA: A multi‐institutional exercise

Alejandra Rangel, Ryan Studinski, Ady Abdellatif, Leigh Conroy, Jenna King, Humza Nusrat iD, Michael Oliver

Journal of Applied Clinical Medical Physics · 2026-08 · 27 (8)

DOI: 10.1002/acm2.70674 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 13.08.2026. Verfasst von Alejandra Rangel und 6 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Department, Trillium Health Partners - Credit Valley Hospital, Mississauga, Ontario, Canada. Hauptthemen laut MEDLINE: Quality Assurance, Health Care, Healthcare Failure Mode and Effect Analysis, Neoplasms, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ontario Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

External beam radiation therapy (RT) involves many complex steps to safely and reliably treat patients. The radiation oncology community has adopted process maps and failure modes and effects analysis (FMEA) to analyze risk using the TG‑100 methodology, but the traditional TG‑100–endorsed risk priority number (RPN) has recognized limitations. An updated risk‑profile introduced by the automotive industry, action priority (AP), offers a severity‑weighted alternative that may better support risk‑based decision‑making in RT.

Purpose

To compare the traditional RPN‐based FMEA approach with the updated AP‐based approach for identifying high‐risk potential failure modes (PFMs) in a representative external beam RT workflow across multiple institutions, and to assess whether the AP framework provides additional value to the radiation oncology community.

Methods

Nine physicists from different cancer centers in Ontario, Canada collaboratively developed and refined a list of 63 PFMs associated with a generic RT workflow requiring data transfer. Each PFM was scored for occurrence (O), severity (S), and detectability (D) using TG‑100 recommendations. Three scoring scenarios were evaluated: average, worst‑case, and a severity‑focused score. High‑risk PFMs were identified using the traditional (top RPN quartile and/or S ≥ 8) and the updated (High or Medium AP categories) FMEA criteria.

Results

Seven high‑risk PFMs were selected using the updated FMEA, and an additional 23 were selected using the traditional FMEA. All PFMs identified by the updated FMEA approach were also selected by the traditional FMEA approach; however, the traditional FMEA approach identified many additional PFMs due to its quartile‑based and severity‐threshold criteria. The updated AP approach appeared less affected by the scoring variability which may reduce the need for consensus discussion, it was generally easier to apply, showed more consistent behaviour across FMEA iterations and provided a clearer action guidance.

Conclusions

The use of the updated AP approach in this study suggests a clearer, severity‐driven, and more streamlined strategy to prioritize high‐risk PFMs, potentially reducing the burden of multi‑institutional FMEA while maintaining consistent identification of critical risks.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quality Assurance, Health CareHealthcare Failure Mode and Effect AnalysisNeoplasmsRadiotherapy Planning, Computer-AssistedHumansRadiotherapy Dosage

Stichwörter der Autor:innen

FMEAaction prioritymulti‑institutional studyquality managementrisk priority number

Autor:innen und Institutionen

  • Medical Physics Department, Trillium Health Partners - Credit Valley Hospital, Mississauga, Ontario, Canada.
  • Department of Physics, Juravinski Hospital and Cancer Centre at Hamilton Health Sciences, Hamilton, Ontario, Canada.
  • Medical Physics Department, R. S. McLaughlin Durham Regional Cancer Centre, Lakeridge Health, Oshawa, Ontario, Canada.
  • Radiation Medicine Program, Princess Margaret Cancer Centre, Toronto, Ontario, Canada.
  • Medical Physics Department, Hudson Regional Cancer Program (North Simcoe Muskoka), Royal Victoria Regional Health Centre, Barrie, Ontario, Canada.
  • Department of Radiation Oncology, Henry Ford Health, Detroit, Michigan, USA.

ORCID-Profile: Humza Nusrat

Förderung

  • Ontario Health

Zitieren

APA

Rangel, A., Studinski, R., Abdellatif, A., Conroy, L., King, J., Nusrat, H., & Oliver, M. (2026). Application of an updated severity‐based FMEA approach to Radiation Therapy QA: A multi‐institutional exercise. Journal of Applied Clinical Medical Physics, 27(8). https://doi.org/10.1002/acm2.70674

Vancouver

Rangel A, Studinski R, Abdellatif A, Conroy L, King J, Nusrat H, et al. Application of an updated severity‐based FMEA approach to Radiation Therapy QA: A multi‐institutional exercise. Journal of Applied Clinical Medical Physics. 2026;27(8). doi:10.1002/acm2.70674

BibTeX
@article{Rangel2026,
  author = {Rangel, Alejandra and Studinski, Ryan and Abdellatif, Ady and Conroy, Leigh and King, Jenna and Nusrat, Humza and Oliver, Michael},
  title = {Application of an updated severity‐based FMEA approach to Radiation Therapy QA: A multi‐institutional exercise},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {8},
  doi = {10.1002/acm2.70674}
}