Dosimetric comparison of FF and FFF modes under non‐ideal bolus‐fitting conditions for post‐mastectomy chest wall irradiations

Long-Gang Gui, Yin Poo

Journal of Applied Clinical Medical Physics · 2026-06 · 27 (6)

DOI: 10.1002/acm2.70653 ↗

At a glance

Vergleichsstudie in Journal of Applied Clinical Medical Physics, published online on 2026-06-08. Written by Long-Gang Gui and 1 co-authors. Institution: School of Electronic Science and Engineering, Nanjing University, Nanjing, China. Major MEDLINE topics: Thoracic Wall, Monte Carlo Method, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted. 19 references. The full text is freely available (open access, CC BY 4.0). Funded by: National Natural Science Foundation of China.

AbsztraktSubjectsBibliográfiai adatokCiteRelated studies

Abstract (original language)

Background

In post‐mastectomy radiotherapy, tissue‐equivalent bolus is routinely used to ensure adequate skin dose. However, achieving perfect contact on curved chest walls is challenging, often resulting in non‐uniform air gaps. With the increasing adoption of Flattening Filter‐Free (FFF) beams, it is crucial to understand how their distinct physical properties—specifically a softer energy spectrum—respond to these low‐density gaps compared to Flattening Filter (FF) beams. Additionally, the accuracy of clinical dose calculation algorithms in these electronic disequilibrium regions remains a subject of concern.

Purpose

This study aimed to quantify the impact of non‐uniform air gaps on skin dose distribution, evaluate the accuracy of Collapsed Cone Convolution (CCC) versus GPU‐accelerated Monte Carlo (MC) algorithms in complex geometries, and determine the specific dosimetric thresholds for selecting FF or FFF beam modes.

Methods

A cylindrical curved phantom (25 cm diameter) was constructed to simulate chest wall anatomy. Four interchangeable modules were designed to create controlled fit conditions: perfect fit, gradient gap (0–6 mm), and uniform gaps (3 and 6 mm). Skin dose measurements were performed using EBT3 films and MOSFET detectors. Dose calculations were conducted in the United Imaging uRT‐TPS using 6 MV FF and FFF beams at gantry angles of 0°, 30°, and 60°.

Results

The presence of gradient gaps substantially increased skin dose inhomogeneity, yielding a heterogeneity index of 0.28 for the FFF beam. The MC algorithm demonstrated high accuracy, agreeing with film measurements within 0.5%. In contrast, the CCC algorithm exhibited systematic underestimation in gap regions, with a mean error of −6.8% and a maximum deviation of −16.5%. Physically, the FFF beam (mean energy 1.23 MeV) was found to be more sensitive to air gaps than the FF beam (1.93 MeV) due to reduced lateral electron scatter. The FFF/FF dose ratio decreased linearly with increasing gap thickness (slope: −2.1%/mm). A clinical threshold was identified at approximately 2.3 mm; beyond this gap thickness, the FF beam exhibited superior dosimetric robustness. Furthermore, oblique incidence at 60° reduced the mean skin dose by approximately 9% compared to perpendicular incidence.

Conclusion

Non‐uniform air gaps significantly degrade skin dose uniformity, with FFF beams being more susceptible to this effect than FF beams. The CCC algorithm shows limitations in these complex interface geometries, whereas the MC algorithm provides a reliable verification tool. Clinically, for regions where air gaps exceed 2.3 mm, the use of FF mode or optimization of bolus fitting is recommended to ensure adequate target coverage.

Subjects

MeSH terms (MEDLINE)

Thoracic WallMonte Carlo MethodPhantoms, ImagingRadiotherapy Planning, Computer-AssistedAlgorithmsMastectomyBreast NeoplasmsHumansRadiotherapy DosageFemaleRadiotherapy, Intensity-ModulatedOrgans at RiskRadiometryParticle Accelerators

Author keywords

Monte Carlo algorithmbolus fitchest wall radiotherapycurved phantomdose heterogeneity index

Authors and institutions

  • School of Electronic Science and Engineering, Nanjing University, Nanjing, China.
  • Radiotherapy Center, Northern Jiangsu People's Hospital Yangzhou, Yangzhou, China.

Funding

  • National Natural Science Foundation of China

Cite

APA

Gui, L. G., & Poo, Y. (2026). Dosimetric comparison of FF and FFF modes under non‐ideal bolus‐fitting conditions for post‐mastectomy chest wall irradiations. Journal of Applied Clinical Medical Physics, 27(6). https://doi.org/10.1002/acm2.70653

Vancouver

Gui LG, Poo Y. Dosimetric comparison of FF and FFF modes under non‐ideal bolus‐fitting conditions for post‐mastectomy chest wall irradiations. Journal of Applied Clinical Medical Physics. 2026;27(6). doi:10.1002/acm2.70653

BibTeX
@article{Gui2026,
  author = {Gui, Long-Gang and Poo, Yin},
  title = {Dosimetric comparison of FF and FFF modes under non‐ideal bolus‐fitting conditions for post‐mastectomy chest wall irradiations},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {6},
  doi = {10.1002/acm2.70653}
}
Start for free →