Surrogate gating strategies for the Elekta Unity MR‐Linac gating system
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.14566 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Samuel D Rusu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Iowa Hospitals and Clinics, Iowa City, Iowa, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Magnetic Resonance Imaging, Particle Accelerators, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
MRI‐guided adaptive radiotherapy can directly monitor the anatomical positioning of the intended target during treatment with no additional imaging dose. Elekta has recently released its comprehensive motion management (CMM) solution that enables automatic radiation beam‐gating on the Unity MR‐Linac. Easily visualized targets that are distinct from the surrounding anatomy can be used to drive automatic gating decisions from the MRI cine imaging. However, poorly visualized targets can compromise the tracking and gating capabilities and may require surrogate tracking structures. This work presents strategies to generate robust tracking surrogates for a variety of treatment sites, enabling a wider application of CMM.
Methods
Surrogate tracking strategies were developed from a cohort of patients treated using the CMM system on the Unity MR‐Linac for treatment sites of the lung, pancreas, liver, and prostate. These sites posed challenging visualization or tracking of the primary target thereby compromising the tracking accuracy. Surrogate structures were developed using site‐specific strategies to improve the imaging textured detail within the tracking volume while avoiding the dynamic overwhelming hypo‐ or hyper‐intense anatomical structures. These surrogate volumes were applied within the anatomical positioning monitoring system as a proxy that drove the CMM gating decisions on the treatment unit.
Results
Robust site‐specific surrogate structures were developed. Surrogate tracking structures for centrally located thoracic targets were created by expanding the target peripherally away from the heart and great vessels and into the lung. Pancreas surrogates required a vertically expanded column intersecting with the inferior liver edge. For the liver and prostate, surrogate structures consisted of a uniform expansion of the target, with liver surrogates intersecting the proximal liver edge or diaphragm while avoiding nearby ribs.
Conclusion
These surrogate strategies have enabled the gating of complex moving targets among different treatment sites at our institution.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Iowa Hospitals and Clinics, Iowa City, Iowa, USA.
ORCID-Profile: Blake R Smith, Joel J St-Aubin
Zitieren
APA
Rusu, S. D., Smith, B. R., St-Aubin, J. J., Shaffer, N., & Hyer, D. E. (2025). Surrogate gating strategies for the Elekta Unity MR‐Linac gating system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14566
Vancouver
Rusu SD, Smith BR, St-Aubin JJ, Shaffer N, Hyer DE. Surrogate gating strategies for the Elekta Unity MR‐Linac gating system. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14566
BibTeX
@article{Rusu2025,
author = {Rusu, Samuel D and Smith, Blake R and St-Aubin, Joel J and Shaffer, Nathan and Hyer, Daniel Ellis},
title = {Surrogate gating strategies for the Elekta Unity MR‐Linac gating system},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.14566}
}