Beyond TG‑43: A PRISMA‐based systematic review on model‐based dose‐calculation algorithms in brachytherapy
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70627 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Dat Tran und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, University of Houston, Houston, Texas, USA. Hauptthemen laut MEDLINE: Brachytherapy, Radiotherapy Planning, Computer-Assisted, Algorithms, Monte Carlo Method. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 69 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background and purpose The AAPM TG‐43 formalism has long served as the clinical standard for brachytherapy dose calculation but assumes a homogeneous water equivalent medium, overlooking limited scattering conditions and tissue heterogeneities. Model‐based dose‐calculation algorithms (MBDCAs), including Monte Carlo (MC) simulations overcome these limitations by accounting for real tissue composition, scatter, and applicator attenuation. This systematic review evaluates TG‐43, MBDCAs/MC methods in terms of dosimetric accuracy, validation strategies, computational feasibility, clinical implementation barriers, and emerging innovations.
Methods
A PRISMA‐guided literature search was conducted using the Scopus database, identifying 284 records, of which 42 full‐text studies met inclusion criteria. Eligible studies compared at least two of the three dose‐calculation approaches (TG‐43, MBDCAs and MCs) in pelvic, breast, or head‐and‐neck brachytherapy. Extracted data encompassed dosimetric discrepancies, validation approaches, computational performance, workflow integration, and enabling technologies.
Results
Across anatomical sites, TG‑43 showed no consistent bias. Its differences from heterogeneity‑aware models depended on tissue composition, scatter conditions, and source geometry. In soft‑tissue regions with minimal heterogeneity, TG‑43 generally overestimated target coverage by about 0.5%–5%. Near low‑density interfaces or in reduced‑scatter configurations, TG‑43 could instead yield lower doses than Monte Carlo or MBDCAs. For OARs, discrepancies were site‑specific: skin dose was often overestimated, while other organs showed smaller or opposite variations. Overall, MBDCAs and Monte Carlo agreed with experimental or benchmark data within roughly 3% and produced more reliable biological metrics. GPU‑accelerated and deep‑learning engines reduced computation times from hours to seconds, shifting remaining challenges toward standardization, commissioning, and QA. Successful clinical adoption relied on TG‑186–aligned validation, staff training, and integrated automated workflows.
Conclusion
Evidence from the systematic review supports a clinical transition toward TG‐186–compliant, heterogeneity‐aware dose‐calculation frameworks. MBDCAs/MC algorithms provide superior dosimetric and radiobiological accuracy and are increasingly compatible with adaptive and biologically guided planning. Broad implementation is now supported by established QA standards and benchmarking datasets, which connects dosimetric precision with patient outcomes. The transition beyond TG‐43 marks a pivotal step toward precision, safety, and personalization in modern brachytherapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physics, University of Houston, Houston, Texas, USA.
- Department of Medical Physics, Faculty of Medicine, Nguyen Tat Thanh University, Ho Chi Minh City, Vietnam.
- Department of Radiation Oncology, UPMC Hillman Cancer Center, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
- Centre for Applied Physics and Radiation Technologies, Sunway University, Bandar Sunway, Malaysia.
- School of Mathematics and Physics, University of Surrey, Guildford, UK.
- Department of Radiation Oncology, University of Toronto, Toronto, Canada.
ORCID-Profile: Dat Tran, Thanh-Tai Duong, Shada Wadi-Ramahi, David Bradley, James C L Chow
Zitieren
APA
Tran, D., Duong, T. T., Wadi-Ramahi, S., Bradley, D., & Chow, J. C. L. (2026). Beyond TG‑43: A PRISMA‐based systematic review on model‐based dose‐calculation algorithms in brachytherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70627
Vancouver
Tran D, Duong TT, Wadi-Ramahi S, Bradley D, Chow JCL. Beyond TG‑43: A PRISMA‐based systematic review on model‐based dose‐calculation algorithms in brachytherapy. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70627
BibTeX
@article{Tran2026,
author = {Tran, Dat and Duong, Thanh-Tai and Wadi-Ramahi, Shada and Bradley, David and Chow, James C L},
title = {Beyond TG‑43: A PRISMA‐based systematic review on model‐based dose‐calculation algorithms in brachytherapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70627}
}
EUVIMED