Simulated online adaptive radiotherapy on C‐arm linacs: Integrating high resolution cone‐beam CT into abdominal stereotactic body radiation therapy

Shiqin Su iD, Song Gao iD, Xiao Liang, Hunter Mehrens, Kevin E Casey iD, Zhiqian H Yu iD, Yusung Kim, Comron J Hassanzadeh, Chad Tang, Phuoc Tho Tran, Deborah Schofield, Peter Balter und 1 weitere

Journal of Applied Clinical Medical Physics · 2026-05 · 27 (5)

DOI: 10.1002/acm2.70615 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 12.05.2026. Verfasst von Shiqin Su und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Radiosurgery, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Conventional cone‐beam computed tomography (CBCT) systems, while providing adequate visualization for positioning, have historically been limited by image truncation, scatter contamination, image noise, geometric artifacts, and unreliable CT number accuracy, which preclude their routine use for accurate dose calculation and re‐planning. Recent advances in CBCT detector technology and image reconstruction algorithms have led to the development of next‐generation CBCT platforms with improved image quality and CT number stability.

Purpose

This study aimed to develop and evaluate an efficient daily adaptation workflow on a standard C‐arm linear accelerator by integrating HyperSight CBCT and plan adaptation in RayStation.

Methods

Image quality metrics and CT number accuracy were evaluated and compared with those of an institutional CT simulator and conventional CBCT. Dose calculation accuracy was validated through measurement. The HyperSight CBCT adaptive workflow used both deformable image registration and rigid image registration for contour propagation, followed by manual revision when necessary. The adapted plan was iteratively optimized in RayStation by using the fine‐tune module. End‐to‐end test was done with a spine phantom and a pelvis phantom. Retrospective evaluation of four abdominal SBRT cases (12 fractions) compared estimated delivered and adapted plans with respect to target coverage, organs at risk (OARs) sparing, and monitor unit (MU) usage.

Results

Improved image quality and CT number accuracy compared to conventional CBCT were observed. The adaptation process required 13 ± 3 min per fraction from HyperSight‐CBCT acquisition to final plan evaluation, excluding time for contour revisions. Across all cases, adapted plans demonstrated improved dosimetric quality relative to delivered plans. Gross tumor volume V 100% and planning target volume V 100% were consistently higher in the adapted plans, with an average increase of 6.3% ( P = 0.004). Doses to OARs, including the small bowel and ipsilateral kidney, were reduced or maintained. Adapted plan MU usage was comparable to reference plans, indicating no increase in treatment complexity.

Conclusions

Integrating HyperSight CBCT with RayStation's streamlined plan adaptation capability provided dosimetric advantages. It improved target coverage and OAR sparing without increasing treatment complexity and represents a practical alternative to Ethos or MR‐Linac systems without the need for specialized delivery platforms.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyRadiosurgeryRadiotherapy Planning, Computer-AssistedPhantoms, ImagingParticle AcceleratorsAbdominal NeoplasmsImage Processing, Computer-AssistedHumansRadiotherapy DosageAlgorithmsRadiotherapy, Intensity-ModulatedRetrospective StudiesOrgans at RiskRadiotherapy, Image-Guided

Stichwörter der Autor:innen

adaptative radiotherapycone‐beam CThyperSightrayStation

Autor:innen und Institutionen

  • Department of Radiation Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Department of GU Radiation Oncology, University of Texas MD Anderson Cancer Center, Houston, Texas, USA.

ORCID-Profile: Shiqin Su, Song Gao, Kevin E Casey, Zhiqian H Yu

Förderung

  • NCI NIH HHS

Zitieren

APA

Su, S., Gao, S., Liang, X., Mehrens, H., Casey, K. E., Yu, Z. H., Kim, Y., Hassanzadeh, C. J., Tang, C., Tran, P. T., Schofield, D., Balter, P., & Han, E. Y. (2026). Simulated online adaptive radiotherapy on C‐arm linacs: Integrating high resolution cone‐beam CT into abdominal stereotactic body radiation therapy. Journal of Applied Clinical Medical Physics, 27(5). https://doi.org/10.1002/acm2.70615

Vancouver

Su S, Gao S, Liang X, Mehrens H, Casey KE, Yu ZH, et al. Simulated online adaptive radiotherapy on C‐arm linacs: Integrating high resolution cone‐beam CT into abdominal stereotactic body radiation therapy. Journal of Applied Clinical Medical Physics. 2026;27(5). doi:10.1002/acm2.70615

BibTeX
@article{Su2026,
  author = {Su, Shiqin and Gao, Song and Liang, Xiao and Mehrens, Hunter and Casey, Kevin E and Yu, Zhiqian H and Kim, Yusung and Hassanzadeh, Comron J and Tang, Chad and Tran, Phuoc Tho and Schofield, Deborah and Balter, Peter and Han, Eun Young},
  title = {Simulated online adaptive radiotherapy on C‐arm linacs: Integrating high resolution cone‐beam CT into abdominal stereotactic body radiation therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {5},
  doi = {10.1002/acm2.70615}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche