A preclinical setup for FLASH radiotherapy beam delivery at isocenter using downstream electron scattering
Journal of Applied Clinical Medical Physics · 2026-04 · 27 (4)
DOI: 10.1002/acm2.70582 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 16.04.2026. Verfasst von Ikechi Ozoemelam und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan, USA. Hauptthemen laut MEDLINE: Particle Accelerators, Electrons, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Preclinical investigations have shown that ultra‐high dose rate (UHDR) radiotherapy (>40 Gy/s) also known as FLASH‐RT, reduces normal tissue toxicity while preserving tumor control efficacy compared to conventional dose rate treatments. Achieving a UHDR‐capable LINAC requires modifications including modifications to electron gun current and bending magnet configurations, use of dedicated energy control boards and use of lower energy scatter foils. However, in the absence of engineering support to implement these changes, it is challenging to realize useful field sizes at ultra‐high dose rates using the upstream scatter foils alone.
Purpose
In this study, we present a configuration to achieve UHDR utilizing downstream electron scattering foils to enable FLASH‐RT at isocenter. By changing the scattering foil position and thickness, we aim to characterize achievable dose rates and dose uniformity for preclinical studies.
Methods
A decommissioned Varian CLINAC 21EX linear accelerator was modified for UHDR irradiation by using the 16 MV beam settings while removing the x‐ray target and flattening filter. Downstream beam scattering was evaluated by testing four lead foil thicknesses (0.28–1.2 mm) at three different source‐to‐foil distances (75.1–90.0 cm), with beam profiles measured using Gafchromic EBT‐XD films at isocenter and evaluated using flatness metrics. The PDDs were also measured to characterize the impact of foils on Bremsstrahlung contribution.
Results
Without a scattering foil, peak dose rates exceeded 200 Gy/s but with poor uniformity (flatness = 22.0%). Standard dual‐foil systems achieved good uniformity (flatness = 6.1%) but yielded non‐UHDR (10–11 Gy/s). With downstream scattering, UHDR (43–135 Gy/s) with improved beam uniformity was achieved. At 75.1 cm source to foil distance, Dmax flatness values at isocenter ranged from 3.5% to 1.2% with increasing foil thickness. However, thicker foils resulted in reduced practical range and increased Bremsstrahlung contribution.
Conclusion
This study demonstrates that downstream electron scattering provides a practical approach for achieving UHDR (43–135 Gy/s) while maintaining acceptable beam uniformity close to the isocenter.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan, USA.
- Department od Radiation Medicine, Oregon Health & Science University, Portland, Oregon, USA.
Zitieren
APA
Ozoemelam, I., Paplanus, R., Pillainayagam, M., Roberts, D. A., & Litzenberg, D. W. (2026). A preclinical setup for FLASH radiotherapy beam delivery at isocenter using downstream electron scattering. Journal of Applied Clinical Medical Physics, 27(4). https://doi.org/10.1002/acm2.70582
Vancouver
Ozoemelam I, Paplanus R, Pillainayagam M, Roberts DA, Litzenberg DW. A preclinical setup for FLASH radiotherapy beam delivery at isocenter using downstream electron scattering. Journal of Applied Clinical Medical Physics. 2026;27(4). doi:10.1002/acm2.70582
BibTeX
@article{Ozoemelam2026,
author = {Ozoemelam, Ikechi and Paplanus, Rowan and Pillainayagam, Michael and Roberts, Donald A and Litzenberg, Dale W},
title = {A preclinical setup for FLASH radiotherapy beam delivery at isocenter using downstream electron scattering},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {4},
doi = {10.1002/acm2.70582}
}
EUVIMED