Electron spatially fractionated radiation therapy using a clinically biocompatible 3D‐printed tungsten‐filament GRID
Journal of Applied Clinical Medical Physics · 2026-04 · 27 (4)
DOI: 10.1002/acm2.70567 ↗At a glance
Journal article in Journal of Applied Clinical Medical Physics, published online on 2026-04-19. Written by Melody Liu and 4 co-authors. Institution: Department of Radiation Oncology, Norris Comprehensive Cancer Center, University of Southern California Keck School of Medicine, Los Angeles, Califor…. Major MEDLINE topics: Tungsten, Phantoms, Imaging, Electrons, Radiotherapy Planning, Computer-Assisted. 26 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Background
Spatially fractionated radiation therapy (SFRT) can enhance tumor control by delivering high‐dose subvolumes while sparing normal tissue and stimulating antitumor immunity. However, its application to superficial disease through electron beam radiation therapy remains limited. Electron SFRT (eSFRT) could serve as a strategy for initial tumor downsizing of ulcerated cutaneous and subcutaneous tumors by delivering a superficial, spatially heterogeneous dose, enabling subsequent conventional electron therapy to enhance local control.
Purpose
To evaluate the dosimetric feasibility of eSFRT using a novel 3D‐printed tungsten‐filament GRID (TS) and compare it to lead sheet (LS) references.
Methods
The TS GRID collimators were 3D‐printed from a tungsten alloy filament (91%–93% tungsten by weight, PLA‐based), with a hexagonal array of 27 apertures (1.5 cm diameter, 2.0 cm spacing, 1 mm thickness) in a 10 × 10 cm 2 sheet. LS GRIDs (1.5 and 3.0 mm thick) were used as reference. GRIDs were placed directly on water and anthropomorphic phantoms. Dosimetric measurements for GRID field properties, including percent depth dose (PDD) and crossbeam plane dose profiles at various depths, were performed in a water tank phantom for 6, 9, and 12 MeV electron beams. Key parameters, such as depths of maximum dose (d max ) and doses at 90% (d90) and 50% (d50), were determined. Peak‐to‐valley dose ratios (PVDRs) were also evaluated and compared with data from an anthropomorphic phantom.
Results
The TS GRID collimator produced highly heterogeneous superficial dose distributions with PVDR > 2.0 at depths > 10 mm for 6–9 MeV, supporting spatially fractionated dose delivery. PVDR reductions on curved phantom surfaces were minimal (
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Radiation Oncology, Norris Comprehensive Cancer Center, University of Southern California Keck School of Medicine, Los Angeles, California, USA.
ORCID profiles: Lauren Lukas
Cite
APA
Liu, M., Xu, Z., Zhang, H., Chau, B., & Lukas, L. (2026). Electron spatially fractionated radiation therapy using a clinically biocompatible 3D‐printed tungsten‐filament GRID. Journal of Applied Clinical Medical Physics, 27(4). https://doi.org/10.1002/acm2.70567
Vancouver
Liu M, Xu Z, Zhang H, Chau B, Lukas L. Electron spatially fractionated radiation therapy using a clinically biocompatible 3D‐printed tungsten‐filament GRID. Journal of Applied Clinical Medical Physics. 2026;27(4). doi:10.1002/acm2.70567
BibTeX
@article{Liu2026,
author = {Liu, Melody and Xu, Zhengzheng and Zhang, Hualin and Chau, Brittney and Lukas, Lauren},
title = {Electron spatially fractionated radiation therapy using a clinically biocompatible 3D‐printed tungsten‐filament GRID},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {4},
doi = {10.1002/acm2.70567}
}
EUVIMED