Development of an optically stimulated luminescence radiometer based on sodium chloride and gold nanoparticles for improving radiation detection in radiography
Journal of Applied Clinical Medical Physics · 2026-05 · 27 (5)
DOI: 10.1002/acm2.70539 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 01.05.2026. Verfasst von Mus'ab S Alkasasbeh und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Allied Medical Sciences, Faculty of Applied Medical Sciences, Al al-Bayt University, Mafraq, Jordan. Hauptthemen laut MEDLINE: Sodium Chloride, Gold, Metal Nanoparticles, Optically Stimulated Luminescence Dosimetry. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deanship of Scientific Research at Imam Mohammad Ibn Saud Islamic University (IMSIU).
Abstract (Originalsprache)
Purpose
To evaluate sodium chloride doped with gold nanoparticles (NaCl + GNPs) as a low‐cost, reusable optically stimulated luminescence (OSL) material for diagnostic radiology, and to benchmark its luminescence performance against the commercial standard, aluminum oxide doped with carbon (Al 2 O 3 :C).
Methods
NaCl + GNPs detectors were fabricated through solution‐based doping, solvent evaporation, pellet pressing, and two‐stage thermal treatment. Structural characterization was verified via UV–Vis spectroscopy. OSL performance was assessed under diagnostic X‐ray energies ranging from 40 to 150 kVp, using Al 2 O 3 :C as the reference. Evaluated parameters included OSL signal linearity, reproducibility, bleaching efficiency, fading behavior, minimum detectable signal (MDS), energy dependence, and mass scaling. Measurements were performed using equal and variable detector masses (20–60 g) across a dose range of approximately 6–60 mGy.
Results
NaCl + GNPs exhibited strong linearity ( R 2 > 0.99) across the tested exposure range and produced reproducible OSL signals over repeated irradiations. The material demonstrated an energy‐dependent increase in luminescence comparable to Al 2 O 3 :C, with the 40 g NaCl + GNPs sample achieving 96–97% of the Al 2 O 3 :C OSL signal at 150 kVp. Increasing detector mass enhanced luminescence output proportionally, with 60 g samples exceeding the reference material at energies above 100 kVp. Bleaching efficiency was notably high, with consistently minimal residual signal, confirming excellent reusability. The MDS was approximately equivalent to 0.4–0.5 mGy in relative terms. However, NaCl + GNPs showed faster fading, with ∼50% signal loss over 9 h, indicating the importance of timely readout or controlled storage.
Conclusions
Gold nanoparticle incorporation enhances the luminescent efficiency and usability of NaCl, producing a material that is low‐cost, scalable, and highly effective in bleaching. Although slightly less sensitive than Al 2 O 3 :C, the NaCl + GNPs detectors demonstrate strong performance, favorable reproducibility, and minimal residual signal, making them a practical option for clinical, educational, and resource‐limited dosimetry applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Allied Medical Sciences, Faculty of Applied Medical Sciences, Al al-Bayt University, Mafraq, Jordan.
- Department of Physics, College of Science, Imam Mohammad Ibn Saud Islamic University (IMSIU), Riyadh, Saudi Arabia.
- Laser and Optoelectronics Engineering Department, Dijlah University, Baghdad, Iraq.
- School of Physics, Universiti Sains Malaysia, Penang, Malaysia.
- Faculty of Education, Universiti Teknologi MARA, Puncak Alam Branch, Bandar Puncak Alam, Selangor, Malaysia.
- Physics Department, College of Science, Al-Hussein Bin Talal University, Ma'an, Jordan.
ORCID-Profile: Mus'ab S Alkasasbeh, Thair Hussein Khazaalah
Förderung
- Deanship of Scientific Research at Imam Mohammad Ibn Saud Islamic University (IMSIU)
Zitieren
APA
Alkasasbeh, M. S., Ibnaouf, K. H., Ahmed, N. M., Abdul Rahman, A., Khazaalah, T. H., Abbas, D. N., Idriss, H., & Alsboul, M. (2026). Development of an optically stimulated luminescence radiometer based on sodium chloride and gold nanoparticles for improving radiation detection in radiography. Journal of Applied Clinical Medical Physics, 27(5). https://doi.org/10.1002/acm2.70539
Vancouver
Alkasasbeh MS, Ibnaouf KH, Ahmed NM, Abdul Rahman A, Khazaalah TH, Abbas DN, et al. Development of an optically stimulated luminescence radiometer based on sodium chloride and gold nanoparticles for improving radiation detection in radiography. Journal of Applied Clinical Medical Physics. 2026;27(5). doi:10.1002/acm2.70539
BibTeX
@article{Alkasasbeh2026,
author = {Alkasasbeh, Mus'ab S and Ibnaouf, Khalid Hassan and Ahmed, Naser M and Abdul Rahman, Azhar and Khazaalah, Thair Hussein and Abbas, Dheyaa Nabeel and Idriss, Hajo and Alsboul, Monther},
title = {Development of an optically stimulated luminescence radiometer based on sodium chloride and gold nanoparticles for improving radiation detection in radiography},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {5},
doi = {10.1002/acm2.70539}
}