Measurement of the volume CT dose index on spiral CT scanning with a real‐time ionization chamber
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70469 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Atsushi Fukuda und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Fukushima, Japan. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, Spiral Computed, Radiometry, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The measurement of computed tomography dose index 100 (), which is feasible only through axial scanning, requires that the clinical spiral protocols be replaced with those for axial scanning. The real‐time ionization chamber detects the integral of radiation dose rate profile, enabling the direct verification of the volume CTDI on spiral CT scanning ().
Purpose
This study aimed to develop a direct measurement technique for and compare its accuracy with that measured using axial scanning () or that displayed on the console ().
Methods
A CTDI phantom with a real‐time ionization chamber was placed on the headrest or examination table. was measured with following parameters: tube voltage = 120 kV, effective mAs = 100, and rotation time = 1.00 s. The parameters for measuring were set identical to those used for axial scanning, except for rotation times = 0.33, 0.50, and 1.00, pitch = 0.35, 0.50, 0.75, 1.00, 1.25, and 1.50, and the scanning range = 15 cm. was extracted from the integral of radiation dose rate profile. was subsequently calculated and compared with and . Finally, was measured for 10 clinical protocols and compared with .
Results
The differences between and , and , and and for the head and body phantoms were all < −2.0%. The differences between and for scans on clinical protocols with and without automatic exposure control were < 11.9% and < 10.5%, respectively; these large differences were observed in the dual‐energy twin‐beam protocol. Excluding this protocol yielded differences between the measurements with and without automatic exposure control of < 2.0% and < −3.9%, respectively.
Conclusions
The results showed an excellent agreement between and , supporting the use of the clinical spiral CT scanning to verify using a real‐time ionization chamber.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Fukushima, Japan.
- Department of Radiological Technology, Faculty of Health Science, Kobe Tokiwa University, Kobe, Hyogo, Japan.
- Department of Radiology, Shiga Psychiatric Medical Center, Kusatsu, Shiga, Japan.
- Department of Medical Technology, Toyama Prefectural Central Hospital, Toyama, Toyama, Japan.
- Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Ishikawa, Japan.
ORCID-Profile: Nao Ichikawa, Ayaka Hirosawa, Kosuke Matsubara
Zitieren
APA
Fukuda, A., Ichikawa, N., Hayashi, T., Hirosawa, A., & Matsubara, K. (2026). Measurement of the volume CT dose index on spiral CT scanning with a real‐time ionization chamber. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70469
Vancouver
Fukuda A, Ichikawa N, Hayashi T, Hirosawa A, Matsubara K. Measurement of the volume CT dose index on spiral CT scanning with a real‐time ionization chamber. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70469
BibTeX
@article{Fukuda2026,
author = {Fukuda, Atsushi and Ichikawa, Nao and Hayashi, Takuma and Hirosawa, Ayaka and Matsubara, Kosuke},
title = {Measurement of the volume CT dose index on spiral CT scanning with a real‐time ionization chamber},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70469}
}
EUVIMED