Influence of scan mode, tilt, and radiation dose on CT radiomic metrics

Neha Yadav, Xiaomeng Lei, Steven Y Cen, Joshua Levy, Kristin Jensen, Bino A Varghese iD

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70462 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Neha Yadav und 5 weiteren Autor:innen. Beteiligte Einrichtung: Keck School of Medicine, University of Southern California, Los Angeles, California, USA. Hauptthemen laut MEDLINE: Radiomics, Phantoms, Imaging, Tomography, X-Ray Computed, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Southern California Dean's Pilot Grant, 2020-2021.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Radiomic features derived from computed tomography (CT) are highly susceptible to variations in acquisition parameters, which can introduce confounding effects in multicenter research and reduce diagnostic accuracy. While the effects of parameters such as scanning mode and dose have been studied, the impact of gantry tilt—despite its routine clinical use—remains underexplored in radiomics literature.

Purpose

To systematically evaluate how scan mode (axial vs. helical), gantry tilt (0° vs. 5°), and radiation dose affect CT‐based radiomic metrics using an anthropomorphic liver phantom containing six 3D‐printed texture inserts, with special emphasis on the novel inclusion of tilt.

Methods

Twelve unique image acquisition configurations were scanned on a GE Revolution Apex CT scanner, with each configuration repeated once. Manual segmentation of volumes of interest (VOIs) was performed, and 93 radiomic features spanning six texture families were extracted using PyRadiomics. First‐order dispersion metrics (standard deviation, interquartile range, and coefficient of variation) were analyzed alongside higher‐order features via regression with heatmap visualization, and repeatable, robust, and calibratable features were identified.

Results

Helical scans without tilt generally exhibited lower first‐order dispersion than axial scans. Introducing a 5° tilt reduced dispersion in axial scans but had inconsistent effects in helical scans, with no coherent trend observed. Radiation dose demonstrated an expected inverse relationship with dispersion metrics. Intraclass correlation coefficient (ICC) analysis revealed that 34% of radiomic metrics exhibited good or excellent repeatability across all trials (ICC ≥ 0.6), but only 13% demonstrated good or excellent robustness, highlighting the sensitivity of radiomic metrics to scanning conditions. Regression analysis yielded 31 metrics (33%) that can be calibrated using their significant linear relationships with the parameters varied in this study, thereby allowing researchers to correct for variations in acquisition settings.

Conclusions

These findings underscore the importance of accounting for acquisition variability—including less frequently examined parameters such as tilt—when designing radiomic studies, selecting robust features, and interpreting results in clinical and multicenter studies. This approach helps distinguish meaningful biological variation from imaging artifacts, thereby improving the reliability of radiomic analysis in personalized medicine.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiomicsPhantoms, ImagingTomography, X-Ray ComputedImage Processing, Computer-AssistedRadiotherapy Planning, Computer-AssistedHumansRadiation DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

image acquisitionphantom studyquality controlradiomic analysis

Autor:innen und Institutionen

  • Keck School of Medicine, University of Southern California, Los Angeles, California, USA.
  • The Phantom Laboratory, Greenwich, New York, USA.
  • Department of Physics and Computational Radiology, Oslo University Hospital, Oslo, Norway.

ORCID-Profile: Bino A Varghese

Förderung

  • University of Southern California Dean's Pilot Grant, 2020-2021

Zitieren

APA

Yadav, N., Lei, X., Cen, S. Y., Levy, J., Jensen, K., & Varghese, B. A. (2026). Influence of scan mode, tilt, and radiation dose on CT radiomic metrics. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70462

Vancouver

Yadav N, Lei X, Cen SY, Levy J, Jensen K, Varghese BA. Influence of scan mode, tilt, and radiation dose on CT radiomic metrics. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70462

BibTeX
@article{Yadav2026,
  author = {Yadav, Neha and Lei, Xiaomeng and Cen, Steven Y and Levy, Joshua and Jensen, Kristin and Varghese, Bino A},
  title = {Influence of scan mode, tilt, and radiation dose on CT radiomic metrics},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70462}
}
Kostenlos starten →